Abbott Afinion ACR (88 pages)
Tags
Languages available:
No exact matches found. Please change your query
Need help?
Do you have a question about the product that is not answered in the manual?
- Receive updates on solutions
- Get answers to your question
Product description:
Frequently Asked Questions
What is the Afinion ACR test used for?
The Afinion ACR test is an in vitro diagnostic tool used to quantitatively determine albumin, creatinine, and the albumin/creatinine ratio (ACR) in human urine. It aids in the early diagnosis of nephropathy and helps identify patients at risk for complications of diabetes or hypertension.
What are the storage conditions for the Afinion ACR test cartridges?
The Afinion ACR test cartridges should be stored refrigerated at 2-8°C (36-46°F). They are stable for up to 3 days at room temperature (20-25°C) if necessary, but it is recommended to store them in the refrigerator.
How should the test cartridge be handled before use?
Before use, the Afinion ACR test cartridge must reach an operating temperature of 20-30°C. Remove it from refrigerated storage and leave it in the unopened foil pouch for at least 15 minutes. Open the pouch just before use and avoid touching the optical reading area.
What is the measuring range for the ACR test?
The ACR test results can be reported in two units: mg/mmol or mg/g. The measuring range for albumin is 5.0-200.0 mg/L, creatinine is 1.5-30.0 mmol/L, and the ACR is 0.1-140.0 mg/mmol (1.0-1225.0 mg/g).
What should be done if the ACR value is outside the measuring range?
If the albumin or creatinine concentration is outside the measuring range, the patient sample can be diluted with saline (0.9% NaCl) up to 4 times (1:4) and retested. The ACR result for the diluted sample is valid, and the original results should be multiplied by the dilution factor.
What are the reference ranges for albumin and ACR?
For 24-hour collection, the reference range for albumin is <30 mg/24h, and for timed collection, it is <20 µg/min. The ACR reference range for spot collection is <3 mg/mmol. Microalbuminuria is defined as 30-300 mg/24h or 20-200 µg/min, and clinical albuminuria is >300 mg/24h or >200 µg/min.
What are the potential interferences for the Afinion ACR test?
No significant interference was observed up to certain concentrations in urine, including acetoacetate, acetone, ascorbic acid, bilirubin, creatine, glucose, and others. However, substances not listed may still interfere with the test and cause false results.
What should be done if the test result is questionable?
If the test result is questionable or if clinical signs are inconsistent with the result, analyze the Afinion ACR Controls and retest the sample using a new test cartridge. If the result is still questionable, confirm it with another method. Regularly analyze control materials to verify the performance of the Afinion Analyzer System.
What is the principle of the Afinion ACR assay?
The Afinion ACR test uses a fully automated immunochemical and enzymatic method to determine albumin, creatinine, and the albumin/creatinine ratio. Albumin is quantified using a solid-phase immunoassay, while creatinine is measured using an enzymatic colorimetric test. The Afinion Analyzer measures the color intensity of the membrane to determine albumin levels.
What is the purpose of the desiccant bag in the test cartridge?
The desiccant bag in the test cartridge contains 1 g of silica gel to absorb moisture. It should not be used in the assay and must be discarded in a suitable container. If the desiccant bag is damaged and particles are found on the test cartridge, the cartridge should not be used.