Product description:
Frequently Asked Questions
How to zero the BMT 964 from the front panel?
To zero the BMT 964 from the front panel, press the binary input for approximately 0.5 seconds with a voltage of +24 VDC between pin 11 (+) and pin 10 (-). This action sets the device to zero display. Ensure the cuvette is cleaned with filtered air or oxygen for at least 10 seconds, including the cleaning time for the inlet hose.
What are the analog output specifications for the BMT 964?
The analog outputs provide a concentration-proportional, isolated voltage signal between 0 and 10 V, updated approximately 25 times per second. The current output delivers a 4-20 mA signal according to IEC 381, where 20 mA corresponds to the maximum concentration. The connected device's input resistance should be greater than 1 kΩ for voltage output and less than 600 Ω (optional: 1350 Ω) for current output.
How to configure the BMT 964 for different units of measurement?
The BMT 964 allows configuration of units for pressure (bar, psi, Torr, MPa), ozone concentration (g/Nm³, ppmv, %wt/wt), and date/time formats. These settings can be adjusted using the front panel buttons or the BMT 964 Link software. The unit display can be changed during operation, but the units are not linearly correlated.
What is the purpose of the serial interface (RS-232) on the BMT 964?
The RS-232 serial interface enables communication with a PC or other automation components. It uses an 8-bit data format with one stop bit and no parity (8N1). The baud rate is adjustable between 2400 and 38400 Baud. The interface provides access to concentration, pressure, and status information, including the warm-up state.
What are the safety precautions for operating the BMT 964?
Ozone is a highly toxic gas, and the BMT 964 measures concentrations that may exceed lethal limits. Safety equipment such as an ozone detector should be used. The device must be connected to a surge protection device within the building installation. The power supply should not exceed 15 A at 120 VAC or 10 A at 240 VAC. Always ensure the device is properly grounded and avoid working on live connections.
How does the BMT 964 handle error conditions?
The BMT 964 has an error relay that can be used to detect faults such as overrange, dirty cuvette, or lamp failure. The error relay can be configured to open or close based on the fault condition. The device also maintains an error log with up to 16 entries, which can be viewed and printed using the BMT 964 Link software.
What are the recommended gas flow rates for the BMT 964?
The recommended gas flow rate for the BMT 964 is between 0.2 and 1 l/min. The response time of the device depends on the flow rate, length, and diameter of the gas tubing. A PTFE tube with dimensions 3x5 mm is recommended to minimize delay. The response time can range from 0.4 to 2 seconds depending on the flow conditions.
How to access the parameter settings on the BMT 964?
To access the parameter settings, navigate to the 'Set Parameters' menu using the SCROLL/SET button for at least two seconds. This menu allows configuration of units, alarm thresholds, communication parameters, and other device settings. A four-digit PIN can be used to protect the settings, with the default PIN set to 0000.
What are the functions of the binary outputs on the BMT 964?
The binary outputs are relay contacts used for error and alarm signals, as well as to control the supply of zero gas during automatic cleaning. The available outputs include Lamp Low, Cuvette Dirty, Low Alarm, High Alarm, and Purge. These outputs can switch up to 28 V and handle a maximum current of 0.5 A. The common terminal can be connected to either ground or the power supply positive.
What is the warm-up time for the BMT 964?
The BMT 964 requires a warm-up time of at least 40 seconds and up to 120 seconds after power is applied. During this time, the device displays the maximum concentration, current pressure, and a warm-up status code. The error relay is set to 'Error' during warm-up, and all other relay contacts are open. The analog outputs provide 10 V or 20 mA during this period.