Product description:

The HemoCue HbA1c 501 is a portable device designed for the rapid and accurate measurement of glycated hemoglobin (HbA1c) in whole blood. It is commonly used in clinical and point-of-care settings to monitor long-term blood glucose control in patients with diabetes. The device operates using a photometric measurement principle and provides results within a few minutes. It features a user-friendly interface, minimal sample volume requirements, and is designed for ease of use by healthcare professionals. The HemoCue HbA1c 501 is intended for in vitro diagnostic use and is suitable for use in laboratories and medical facilities requiring quick and reliable HbA1c testing.

Frequently Asked Questions

How does the HemoCue HbA1c 501 Analyzer separate glycated hemoglobin?

The HemoCue HbA1c 501 Analyzer uses a boronate affinity assay to separate the glycated hemoglobin fraction from the nonglycated fraction. The boronate resin binds to the cis-diols of glycated hemoglobin, allowing for precise measurement.

What is the operating principle of the HemoCue HbA1c 501 Analyzer?

The HemoCue HbA1c 501 Analyzer uses a boronate affinity assay to measure hemoglobin A1c (HbA1c) in whole blood. The blood sample is lysed, and the hemoglobin is measured photometrically using an optical sensor composed of a LED and a PD (Photo Diode).

What is the formula for calculating HbA1c percentage?

The HbA1c percentage is calculated using the formula: HbA1c % = A x [HbA1c / Total Hemoglobin] x 100 + B, where A and B are slope and intercept factors used to correct the value for the calibration standard of NGSP.

How long does the HemoCue HbA1c 501 Analyzer take to warm up?

The HemoCue HbA1c 501 Analyzer takes approximately 5 minutes to warm up, depending on the ambient temperature. During this time, it performs a hardware functionality test to ensure the internal optics and mechanical system are operating correctly.

What types of blood samples can be used for HbA1c testing?

The HemoCue HbA1c 501 Analyzer can process both capillary blood samples and venous whole blood collected using K2EDTA, lithium heparin, sodium citrate, or sodium fluoride/oxalate as an anticoagulant. The sample volume required is 4 µL.

How do you insert the reagent pack into the analyzer?

To insert the reagent pack, open the analyzer lid and place the reagent pack with the blood sample into the cartridge compartment. Ensure the barcode is facing left and gently push it in until a snap is heard, confirming proper placement.

What should you do if the HbA1c test result is outside the normal range?

If the HbA1c test result is outside the normal range (e.g., >14% or <4%), repeat the test to confirm the result. If the second result is also outside the range, contact your local distributor for further assistance.

How do you perform a system check with the HemoCue HbA1c 501 Analyzer?

To perform a system check, use the Daily Check Cartridge or Monthly Check Cartridge. These cartridges screen the optical and operating systems of the analyzer, ensuring accurate and reliable results. Follow the on-screen instructions to complete the check.

What are the storage conditions for the HemoCue HbA1c 501 Analyzer?

The HemoCue HbA1c 501 Analyzer should be stored at 10-35°C (50-95°F). Test Cartridges must be stored at 2-32°C (36-90°F) with a relative humidity of 10-90% until the expiry date printed on the package.

What are the safety precautions when using the HemoCue HbA1c 501 Analyzer?

Always handle the analyzer with care to avoid damage to internal components. Use gloves when handling blood samples to prevent biological hazards. Dispose of used test cartridges in a sturdy container and follow local regulations for proper disposal and disinfection.