Product description:
Frequently Asked Questions
How do I install the batteries in the SNAP Pro Analyzer?
Open the battery compartment on the back of the analyzer and insert two batteries, negative end first. Ensure they are fully seated for proper operation.
What steps should I follow to connect the power adapter?
Remove the protective insert while sliding the locking key toward the cord. Insert the adapter top-side first and snap it into place. Connect the power cord to an appropriate electrical outlet, ensuring a steady yellow status light appears.
How long does it take to charge the analyzer fully?
It takes up to 8 hours for the analyzer to fully charge. The status light will turn green when charging is complete. You can use the device while it's charging, but avoid overcharging.
How do I power on the SNAP Pro Analyzer?
Press the gray Power button at the top of the back to start the analyzer. Follow the on-screen instructions for setup, including selecting your location and date/time format.
What should I do if there are light spots or unexpected colors in test results?
If you notice very light spots or unusual color development in the results window, you may need to confirm some or all of the test's outcomes. Highlighted yellow results require visual confirmation using the provided instructions.
How do I clean the case and screen of the analyzer?
Use a lint-free cloth dampened with 70% isopropyl alcohol, mild soap, or a 10% bleach solution to wipe down the case and screen. Avoid cleaning vents under the power button or ports.
How do I clean internal components of the analyzer?
Power off the device and disconnect all cords before removing the base. Clean the interior using a lint-free wipe with 70% isopropyl alcohol, soap, or bleach solution. Ensure proper reattachment to maintain performance.
What do the SNAP Pro icons indicate?
The wireless signal status shows five gray bars for best connection and red bars if no signal exists. A green battery icon means full charge, while a yellow one indicates low power or need for confirmation of results.
How can I confirm test results manually on the SNAP Pro Analyzer?
If highlighted results require verification, tap the plus icon to mark them as positive or minus icon if negative. After confirming all necessary outcomes, select the Check Mark icon to save and send results.
What should I do if there is no communication with the IDEXX VetLab Station?
Check that your analyzer is properly connected via a wired link or wireless network. If issues persist, ensure the router is functioning correctly and restart both devices to reestablish connectivity.