WIKA T16 (68 pages)
Tags
Languages available:
No exact matches found. Please change your query
Need help?
Do you have a question about the product that is not answered in the manual?
- Receive updates on solutions
- Get answers to your question
Product description:
Frequently Asked Questions
What is the model number of the digital temperature transmitter?
The digital temperature transmitter is model T16. It is available in head mounting version T16.H and rail mounting version T16.R.
What are the operating and instruction pages for the T16 model?
The English operating instructions for model T16 cover pages 3 to 36, while the German version covers pages 37 to 65. These pages include general information, design, safety, and technical specifications.
What safety precautions should be taken when using the T16 temperature transmitter?
Before starting any work, read the operating instructions carefully. Ensure that all safety instructions are followed, and that the device is used only for its intended purpose. Skilled personnel must handle the device, and proper labeling and safety marks must be maintained.
What are the safety-related characteristic values for the T16 models?
The safety-related characteristic values for models T16.x-AI and T16.x-AC are specified in the technical documentation. Similarly, the values for models T16.x-AN and T16.x-AE are also detailed in the specifications section. These values ensure compliance with safety standards and proper operation in hazardous areas.
How is the T16 temperature transmitter configured?
The T16 temperature transmitter can be configured via a PC using the WIKAsoft-TT configuration software. It also supports configurable monitoring functionality, and can be programmed using a programming unit model PU-548. The configuration process includes setting up the device for specific applications and monitoring parameters.
What are the technical specifications of the T16 temperature transmitter?
The T16 temperature transmitter converts thermoelectric voltage into a proportional current signal (4 ... 20 mA). It supports Pt100/3 sensors with a temperature range of -150 ... +180 °C. The device is designed for use in hazardous areas and meets explosion protection standards such as ATEX and IECEx.
What are the approval standards for the T16 temperature transmitter?
The T16 temperature transmitter is approved for use in hazardous areas under ATEX and IECEx standards. Different models, such as T16.x-AC, T16.x-AI, T16.x-AN, and T16.x-AE, have specific approvals for different explosion protection categories. These approvals ensure safe operation in environments with potential explosive hazards.
What are the dimensions of the T16 temperature transmitter?
The T16 temperature transmitter is available in two mounting versions: head mounting (T16.H) and rail mounting (T16.R). The dimensions are specified in millimeters, and the exact measurements are provided in the design and function section of the operating instructions.
What is the purpose of the configuration software WIKAsoft-TT?
The WIKAsoft-TT configuration software is used to configure the T16 temperature transmitter. It allows users to set up the device for specific applications, monitor its performance, and perform fault diagnosis. The software also supports configuring multiple instruments identically, ensuring consistency across installations.
What are the requirements for personnel working with the T16 temperature transmitter?
Only skilled electrical personnel should handle the T16 temperature transmitter. These personnel must have the necessary qualifications, including knowledge of ignition protection types, regulations for hazardous areas, and current safety standards. They must also be trained to recognize and avoid potential hazards during installation and operation.