NSK Primado2 (152 pages)
Tags
Languages available:
No exact matches found. Please change your query
Need help?
Do you have a question about the product that is not answered in the manual?
- Receive updates on solutions
- Get answers to your question
Product description:
Frequently Asked Questions
What is the purpose of this reprocessing manual?
This manual provides detailed instructions for reprocessing the Primado2 surgical system to ensure safety, functionality, and compliance with ISO 17664 standards. It covers cleaning, checking, and sterilization procedures for all components.
What are the cutting accessories called in this manual?
In this manual, the terms 'Bur', 'Drill', 'Blade', and 'Rasp' are collectively referred to as 'Cutting Accessory(ies)'. These are single-use items and must not be reused or reprocessed.
What should I do before using the Primado2 system?
Before using the Primado2 system, you should carefully read this manual to understand the correct reprocessing procedures. This ensures the continued safe and effective use of the attachments and other components.
What are the risks of reusing the cutting accessory?
Reusing the cutting accessory can lead to infection, personal injury due to damaged accessories, and system defects caused by damaged or deformed parts. It is strictly prohibited to reuse or reprocess these items.
How should I clean the control unit?
To clean the control unit, use a cloth moistened with alcohol disinfectant (e.g., Meliseptol® rapid). Avoid submerging it in water or using chlorinated cleaners, as these can damage the unit. Ensure no water enters the connector area.
What is the recommended cleaning procedure for the motor?
For manual cleaning of the motor, use a soft bristle brush with neutral enzymatic detergent under running tap water. Rinse thoroughly with distilled or deionized water, and dry with a clean cloth or air gun. Avoid using any lubricants on the motor.
What should I do if there is residual contamination after cleaning?
If there is residual contamination after cleaning, repeat the pre-cleaning through drying procedures until all contamination is removed. Ensure the product is visually inspected and meets the required standards before proceeding to checking or sterilization.
What are the steps for checking the motor after cleaning?
After cleaning, check for any abnormalities or damage on the motor's outer surface. Ensure the motor can be securely connected to the control unit. Test the pedal for abnormal vibration, noise, or heating. Verify that the rotational speed reaches the maximum setting on the control unit.
What is the recommended disinfectant for cleaning the control unit?
The recommended disinfectant for cleaning the control unit is alcohol-based, such as ethanol or ethanol with isopropanol added (e.g., Meliseptol® rapid). Avoid using chlorinated cleaners, benzene, or thinner, as they can damage the unit.
What should I do after completing the reprocessing steps?
After completing the reprocessing steps, proceed to the storage section (see '11 Storage') to ensure the device is properly stored. If the reprocessing was done manually, follow the packaging and sterilization procedures outlined in section 10 to prepare the device for use.
Related products from NSK
prophy-mate neo
User manual 82 pages
Presto II
User manual 21 pages
Surgic XT Plus
User manual 97 pages
PRESTO AQUA II
User manual 11 pages
PRESTO AQUA II
User manual 43 pages
VarioSurg
User manual 122 pages
Osseo 100
User manual 26 pages
Osseo 100
User manual 36 pages
Osseo 100
User manual 84 pages
Ti-Max S970L
User manual 36 pages
Ti-Max S970
User manual 36 pages
Ti-Max S970SL
User manual 36 pages
Ti-Max S970KL
User manual 36 pages
Ti-Max X600L
User manual 47 pages
Ti-Max X700L
User manual 47 pages
Ti-Max X500
User manual 47 pages
Ti-Max X500L
User manual 47 pages
Ti-Max X600
User manual 47 pages
Ti-Max X700
User manual 47 pages
DynaLED M205LG M4
User manual 54 pages