Product description:

The Current Clamp Greenlee CM-950 is a precision measuring device designed for the accurate detection and measurement of electrical currents in conductors without the need for direct contact. It features a robust construction and a clear display for easy reading. The device is suitable for use in a variety of electrical applications, including fault finding, load testing, and energy monitoring. Its ergonomic design ensures comfortable handling during extended use. The clamp is compatible with a wide range of conductor sizes and is equipped with a high-precision sensor for reliable measurements.

Frequently Asked Questions

How do I turn on the backlight?

Press the backlight button to illuminate the display. Press it again to turn off. The backlight will automatically shut off after approximately 1 minute of inactivity.

What does the HOLD button do?

Press the HOLD button momentarily to freeze the current display value. Press it again to return to normal operation mode.

How long does the unit stay on when not in use?

The unit automatically turns off after approximately 30 minutes of inactivity. This feature can be disabled by pressing the PEAK, MIN MAX, or ZERO button while turning the meter on.

What is the purpose of the MIN MAX button?

The MIN MAX button allows you to record and display the minimum and maximum values of the input. Press it to cycle through display modes, and hold it to exit this mode.

What does the PEAK button do?

The PEAK button records and displays the minimum and maximum peak values of AC inputs. Press it to cycle through display modes, and hold it to exit this mode. A brief 'CAL' message appears as internal circuitry resets.

How do I zero the meter for DC current measurements?

Press the ZERO button momentarily to compensate for residual magnetism in the jaw. This will display the relative measurement. Press again to view the uncompensated measurement. Hold the button to clear the stored offset and return to uncompensated mode.

What is the difference between AC RMS and true RMS measurements?

AC RMS measurements using an average-responding meter are accurate for pure sine waves but less so for other waveforms. True RMS meters, like the CM-950, provide accurate readings for any waveform, including square, triangle, or harmonic signals, making them more versatile for various applications.

How do I measure DC current?

To measure DC current, set the selector to the appropriate range (start with 600 A), clamp the jaw around the conductor, and press the ZERO button to compensate for residual magnetism. Ensure the jaw is fully closed and the wire is centered for accurate readings.

What does the continuity test function do?

The continuity test function emits a tone if the measured resistance is less than approximately 30 ohms, indicating a continuous circuit. This is useful for verifying connections and checking for breaks in a circuit.

How do I connect the test leads for resistance measurement?

For resistance measurement, connect the red lead to the V-Ω terminal and the black lead to the COM terminal. Ensure the circuit is powered off and the meter is set to the appropriate range for accurate readings.