Product description:

The Struers Unitom-2 is a precision instrument designed for the preparation of metallographic samples. It features a robust construction and advanced automation to ensure consistent and high-quality results. The device allows for accurate control of grinding and polishing parameters, making it suitable for a wide range of materials. Its modular design enables easy maintenance and adaptation to different sample preparation requirements. The Unitom-2 is commonly used in laboratories and industrial settings where precise surface preparation is essential for material analysis and quality control.

Frequently Asked Questions

How do I check the contents of the packing box?

In the packing box, you should find a fork spanner, grease gun, cover plugs, drain angles, outlet hose, hose clamps, and a set of instruction manuals. Ensure all parts are present before proceeding.

How do I unpack the Unitom-2?

Remove the wooden block holding the spindle using a 17 mm spanner. Insert the two cover plugs into the threaded holes. Ensure all components are accounted for and in good condition.

How should I place the Unitom-2?

Place the Unitom-2 on an optional Struers Table Unit or a table that can support a minimum weight of 200 kg. Adjust the table height using the screws in the legs. Ensure the table and Unitom-2 are level.

How do I adjust the height of the Unitom-2?

Loosen the upper 13 mm nut under the corner screw, adjust the lower 8 mm nut with a spanner, and fasten the upper nut again. Repeat for each corner until the Unitom-2 is level.

What is the noise level of the Unitom-2?

The Unitom-2 produces approximately 76 dB (A) at idle, measured at a distance of 1.0 m from the machine. This is a standard noise level for similar equipment.

How do I connect the cooling unit to the Unitom-2?

Plug the Cooli control unit’s communication cable into the Unitom-2’s control socket. Connect the quick coupling on the cooling fluid inlet tube to the cooling unit’s pump outlet. Ensure the outlet hose slopes downward and is properly clamped.

What should I do before connecting the cooling unit?

Follow the instructions in the Cooli Instruction Manual to prepare the cooling unit for use. Ensure the mains voltage matches the voltage stated on the machine’s type plate before connecting.

How do I check the direction of the cut-off wheel?

Verify that the cut-off wheel rotates in the direction indicated on the label attached to the guard. If the direction is incorrect, switch two of the phases or adjust the wiring accordingly.

What safety precautions should I follow when using the Unitom-2?

Always wear safety goggles when using the flushing hose, use safety shoes for heavy workpieces, and ensure the machine is placed on a stable surface. Avoid using damaged cut-off wheels and follow all safety regulations for handling cooling fluid.

What should I do if the Unitom-2 is not level?

Adjust the height of the Unitom-2 by loosening the upper 13 mm nut, adjusting the lower 8 mm nut with a spanner, and tightening the upper nut again. Repeat for each corner until the machine is level.