Product description:
Frequently Asked Questions
How to connect the power supply to the docking station?
The power supply is connected to the docking station using the terminal block. Terminal 1 is for the positive voltage, terminal 2 is for the negative voltage, and terminal 5 is for the ground connection. Ensure all connections are secure and properly insulated.
What is the terminal assignment for the 4-20 mA signal?
The 4-20 mA signal is connected to terminal 3. This terminal provides the analog signal output for the docking station. Ensure the signal is properly grounded and connected to the appropriate device.
How to connect the relay to the docking station?
The relay is connected to the relay terminal block. Terminal 1 is for the fault signal, terminal 2 is for the alarm signal, and terminal 3 is for the common connection. The relay is factory-set to normally open (NO) and common (COM) configuration.
What are the wiring requirements for the fieldbus connection?
For the fieldbus connection, use shielded cables. The positive voltage is connected to terminal PWR+, the negative voltage to terminal PWR-, and the shield to the SCR terminal. Ensure the cables are properly shielded and grounded.
How to connect the Modbus RTU interface?
The Modbus RTU interface is connected using the terminal block. Terminal FB+ is for signal line A, terminal FB- is for signal line B, and the ground is connected to the COM terminal. Ensure the shield is properly grounded and the connections are secure.
What are the minimum wire sizes for the connections?
The minimum wire size for the Polytron SE EX sensing heads is 0.7 mm² (AWG 21). For relay connections, use 1 mm² (AWG 18). For daisy-chaining gas transmitters, use 2 flexible wires with a size of 1 mm², crimped together with a ferrule within a cage clamp.
What is the maximum distance between the gas detector and the sensors?
For remote installations, the maximum permissible distance between the gas detector and the sensors is 30 meters (100 feet). Ensure the cables are properly shielded and the connections are secure to maintain signal integrity.
How to mount the docking station on a flat surface?
To mount the docking station on a flat surface, use a drilling template (part number 45 44 299) and attach it with 4 screws (M6 x 10). Ensure the mounting is secure and the docking station is properly aligned with the surface.
What are the installation requirements for hazardous areas?
The docking station is intended for installation in hazardous areas of zone 1, 2, or 21, 22 according to device categories 2G, 3G, or 2D, 3D. Ensure all wiring and connections comply with the relevant national regulations for explosive areas.
How to secure the gas detector to the docking station?
The gas detector is secured to the docking station using 4 screws with a torque of 8 Nm (70 LB IN). Ensure the screws are properly tightened and the connection is secure to prevent damage during operation.