GCE MEDIMETER (124 pages)
Tags
Languages available:
No exact matches found. Please change your query
Need help?
Do you have a question about the product that is not answered in the manual?
- Receive updates on solutions
- Get answers to your question
Product description:
Frequently Asked Questions
What gases is the flowmeter suitable for?
The flowmeter is intended for the administration of medical gases such as oxygen (O2), medical air, and nitrous oxide (N2O). It is not suitable for use with air or nitrogen to drive surgical tools.
What are the operating conditions for the flowmeter?
The flowmeter should be used in environments with temperatures between -20°C and +60°C, humidity levels between 10% and 100%, and atmospheric pressure between 600 and 1200 mbar. Avoid extreme temperatures during storage.
How do you adjust the flow rate?
To adjust the flow rate, turn the handwheel. Turning it anticlockwise increases the flow, while turning it clockwise decreases the flow. Ensure not to apply excessive torque when reaching the end position.
What is the purpose of the float in the flowmeter?
The float inside the calibrated tube indicates the flow rate. When the flow is constant, the float stabilizes at a specific position on the scale, allowing the user to read the flow value accurately.
How should the flowmeter be cleaned?
Clean the flowmeter with a soft cloth dampened in oil-free, oxygen-compatible soap water and rinse with clean water. Avoid using ammonia-based cleaners, immersion in liquids, or exposure to high temperatures.
What is the maximum lifespan of the flowmeter?
The maximum lifespan of the flowmeter is 10 years. At the end of its life, the product must be removed from service and disposed of properly, following environmental and safety regulations.
How do you perform a leakage test?
To perform a leakage test, connect the flowmeter to a compressed air source (max. 5 bar), immerse the hose end in water, and observe for bubbles. No more than 3 bubbles per minute should be visible during the test.
What parts can be replaced by the user?
Users can replace certain parts such as caps, stickers, hose nipples (including O-rings), and handwheels. However, all repairs and replacements should be done by authorized personnel to ensure safety and proper function.
What should be done before using the flowmeter for the first time?
Before initial use, ensure the product is in its original packaging and check for any visible damage or contamination. Perform a visual inspection and functional test to confirm the flowmeter is in good working condition.
What are the safety requirements for storage and transport?
Store and transport the flowmeter away from heat sources, flammable materials, oil, grease, water, and dust. Maintain temperatures between -30°C and +60°C during transport and avoid exposing the product to high temperatures or liquids.