What is the recommended cleaning method for HiFocus Electrode Forceps?
The HiFocus Electrode Forceps can be cleaned using either manual or automatic methods. For manual cleaning, submerge the tools in alkaline detergent, brush to remove debris, rinse with warm water, and sonicate in an ultrasonic unit. For automatic cleaning, use a washer-disinfector with specified temperature and recirculation times.
What temperature should be used during cleaning?
During cleaning, the water temperature should be warm, typically around 65.5°C (150°F) for the washer-disinfector. For thermal disinfection, the temperature should be at least 90°C (194°F) to ensure proper disinfection.
What should be done if soil remains after cleaning?
If visual soil is present after cleaning, repeat the entire cleaning process. This includes rinsing, brushing, sonication, and rinsing with Reverse Osmosis (RO) or Deionization (DI) water to ensure all contaminants are removed.
What disinfection methods are recommended for HiFocus Electrode Forceps?
Two disinfection methods are recommended: chemical and thermal. Chemical disinfection uses Cidex® 2.4% Glutaraldehyde, while thermal disinfection is performed using a washer-disinfector at 90°C (194°F) as part of an automated cleaning cycle.
Can thermal disinfection be used alone?
Thermal disinfection cannot be used alone; it must be performed in connection with automated cleaning. The disinfection step should be applied after rinsing and prior to the drying step in the washer-disinfector cycle.
What is the recommended chemical disinfectant for HiFocus Electrode Forceps?
The recommended chemical disinfectant is Cidex® 2.4% Glutaraldehyde. The disinfectant should be prepared according to the manufacturer’s instructions, and the concentration should be verified using Cidex 2.4 test strips before use.
How long should the tools be submerged in the disinfectant?
The tools should be fully submerged in the disinfectant solution for 60 minutes at 25°C ± 2°C (77°F ± 3.6°F). After this time, the tools should be rinsed with Deionization (DI) water to remove any residual disinfectant.
What should be done after chemical disinfection?
After chemical disinfection, the tools should be rinsed with Deionization (DI) water for 30 minutes. Use a 10 ml (0.3 oz.) syringe to force DI water into cracks and grooves six times to ensure thorough rinsing.
What is the recommended packaging before sterilization?
After cleaning and disinfection, the tools should be packaged in a manner compliant with ISO 11607-1: 2006 standards. This ensures proper protection and prevents contamination before the sterilization process.
What should be done if multiple tools are being sonicated?
If multiple tools are being sonicated, they should not be in contact with each other during the process to avoid damage or cross-contamination. Ensure each tool is properly separated and submerged in the ultrasonic unit.