Product description:

The Level Electrodes FHC microTargeting are designed for precise and reliable measurement in various industrial and scientific applications. These electrodes feature a microTargeting system that ensures accurate contact with the surface being measured. They are constructed from high-quality materials to ensure durability and long-term performance. The design allows for easy integration into different measurement setups and supports a wide range of operating conditions. Their compact size and robust construction make them suitable for use in challenging environments.

Frequently Asked Questions

How should microTargeting™ Electrodes be handled to avoid damage?

Extreme care must be used in handling the microTargeting™ Electrodes. The tips are extremely small and delicate. The electrode should not be retracted back into the macro tube by more than 3 cm, otherwise the tip may be damaged.

What is the recommended method for sterilizing microTargeting™ Electrodes?

FHC has validated and recommends the following steam sterilization parameters: Prevacuum wrapped (in 2 layers of 1-ply polypropylene wrap), preconditioning pulses: 3, exposure time: 4 minutes at 132°C (270°F), minimum dry time: 30 minutes. The electrodes should be removed from the package prior to sterilization, and FHC recommends use of the FHC sterilization tray.

What are the temperature limits for safe use of microTargeting™ Electrodes?

The medical device can be safely exposed to temperatures ranging from 0°C to 40°C (32°F to 104°F).

What is the maximum inside diameter of the insertion tube for microTargeting™ Electrodes?

microTargeting™ Electrodes must only be inserted through a rigid insertion tube having a maximum inside diameter no more than 100 microns (0.004") larger than the diameter of the electrode.

What is the recommended impedance for microTargeting™ Electrodes?

The impedance measured in vivo should be lower than the nominal tip impedance specified. Discontinue use of the microTargeting electrode if the impedance result is higher or lower than is acceptable.

What are the recommended parameters for macrostimulation using microTargeting™ Electrodes?

Balanced, biphasic square pulses with the following parameters are recommended: Frequency ≤ 300 Hz, Pulse Duration 100-200µs, Train Duration ≤ 5 sec., Current Amplitude ≤ 10mA. Stimulus amplitude should never exceed the recommended guidelines provided by the manufacturer of the stimulus generator.

What should be done if a suitable target location is not found during use?

Additional tracks may be indicated if a suitable target location is not found. Retract the electrode by first disconnecting the patient leads, then retracting the microelectrode tip up into the macro tube, and then retracting the entire microTargeting™ Electrode from the insertion tube.

What is the maximum number of microelectrodes included in the KTF1001 kit?

The KTF1001 kit includes 5 microelectrodes, 5 microelectrode insertion tubes/stylettes, 5 insertion guide tubes, 1 microelectrode lead cable, 1 sterile drape sleeve, 1 lead insertion tube/stylette, and 1 microelectrode spacer tube.

What is the total extension length to the tip for the FC2001 microTargeting™ Electrode?

The total extension length to the tip for the FC2001 microTargeting™ Electrode is 266.4mm.

What is the maximum diameter of the FC4000 microTargeting™ Electrode?

The maximum diameter of the FC4000 microTargeting™ Electrode is 0.56mm.