Product description:
Frequently Asked Questions
How to connect the light guide?
To connect the light guide, first turn the fixing screw counterclockwise to open the light guide socket. Insert the light guide until it clicks into place, then turn the screw clockwise to secure it. Ensure the guide's lead-in fits into one of the two guide slots.
How to turn the device on and off?
Use the main switch to turn the KL 1500 HAL on and off. When the device is on, the LCD display will light up. To disconnect the power, pull the power plug. The device is equipped with a soft start to protect the halogen lamp from high inrush current.
How to adjust the light intensity?
Rotate the light intensity regulator to adjust brightness. The LCD display shows the intensity as a percentage between 0% and 100%. Alternatively, you can view it in Kelvin by pressing the regulator once. Note that this is an approximate value and not a precise measurement.
What does the 'No Light Guide' error mean?
The 'No Light Guide' error appears in the LCD display if no light guide is inserted in the socket while the device is on. This prevents operation without a guide, avoiding accidental exposure to intense light. Ensure the guide is properly inserted before using the device.
How to reset the total operating time?
To reset the total operating time, navigate to the 'Reset' option in the menu settings. Confirm by pressing briefly (a checkmark appears in the menu field). Then select 'Menu' and 'Exit' to save the changes. This resets the lamp usage time after a lamp replacement.
How to use the additional optics?
Use the swivel lever to adjust the additional optics for even illumination. When using light guides with a smaller beam diameter than 9 mm, the swivel ensures uniform lighting. For imaging or focusing systems, extend the optics fully to achieve consistent illumination.
How to install a filter?
Before inserting the filter, ensure the filter slider is cooled down. Wear clean gloves to avoid contamination. Pull the slider completely out and rotate it to make the markings readable. Insert the filter from the top right behind the two black retaining tabs. Press gently from the top right to secure the filter in place.
What safety precautions should be taken?
Avoid looking directly into the light exit or the light guide during operation to prevent eye injury. Ensure the light guide and socket are never covered, as this poses a fire hazard. Keep the device in a dry area and away from explosive environments. Always use the correct voltage as specified on the type plate.
How to handle heat-sensitive materials?
When illuminating heat-sensitive or flammable materials, ensure the light guide distance and lamp brightness are set to avoid thermal damage. Keep all cooling vents clear, as the built-in thermostat will reduce light intensity if cooling is insufficient. Maintain a safe distance of at least 10 cm from heat-sensitive materials.
What are the lamp life indicators?
The display shows three different values based on light intensity, indicating the approximate lamp life: ECO (> 1500 hours), STANDARD (150–1500 hours), and BOOST (50–150 hours). These values help estimate the remaining lifespan of the halogen lamp under different operating conditions.
Related products from SCHOTT
KL 2500 LED
User manual 56 pages
KL 2500 LED
User manual 56 pages
KL 2500 LED
User manual 62 pages
KL 1500 LED
User manual 40 pages
KL 300 LED
User manual 44 pages
KL 300 LED
User manual 47 pages
KL 2500 LCD
User manual 46 pages
KL 1500 H
User manual 47 pages
KL 1500 H
User manual 84 pages
EasyLED
User manual 54 pages
EasyLED
User manual 64 pages
KL 1600 LED
User manual 56 pages
KL 1600 LED
User manual 67 pages