Product description:

The Aqualytic AL100 is a measuring instrument designed for precise analysis of water quality parameters. It is commonly used in environmental monitoring, industrial processes, and laboratory settings. The device features a user-friendly interface and advanced sensor technology for accurate and reliable measurements. It supports various measurement modes and can be easily integrated into existing systems. The AL100 is built to withstand challenging environmental conditions, ensuring long-term durability and consistent performance.

Frequently Asked Questions

How to perform a zero calibration?

To perform a zero calibration, fill the cuvette with clean water up to the 10-ml mark, close it with the lid, and place it in the measurement chamber. Press the [ZERO/TEST] button. The method symbol will blink for approximately 8 seconds. After completing the zero calibration, remove the cuvette from the measurement chamber, add the reagent, and wait for the characteristic color to develop. Close the cuvette again and place it back in the measurement chamber.

How to read stored data?

To read stored data, press and hold the [!] button for more than 4 seconds while the device is on. This will directly take you to the memory menu where you can view previously saved results.

How to perform a test after zero calibration?

After completing the zero calibration, add the reagents to the sample, close the cuvette, and place it back in the measurement chamber. Press the [ZERO/TEST] button. The method symbol will blink for approximately 3 seconds, and the result will appear on the display.

How to handle a reaction time delay?

For methods with a reaction time, you can optionally enable a countdown function. Press and hold the [!] button, then press [ZERO/TEST]. Release the [!] button to start the countdown. After the countdown completes, the measurement will start automatically. The countdown can be stopped by pressing [ZERO/TEST], and the measurement will begin immediately.

What are the measurement tolerances for ammonium?

The measurement tolerances for ammonium are ±0.05 mg/l N. This ensures accurate results within the specified range when following the correct procedure.

How to prepare the sample for ammonium analysis using tablets?

For ammonium analysis using tablets, add 10 ml of the sample to a clean 24-mm cuvette and perform a zero calibration. Add an AMMONIA No. 1 tablet directly from the foil to the 10-ml sample and crush it with a clean stirrer. Add an AMMONIA No. 2 tablet to the same sample and crush it again. Close the cuvette with the lid and mix by inverting until the tablets dissolve.

How to handle the cuvette during measurement?

Ensure the cuvette is clean, dry, and free from fingerprints or water droplets on the light transmission areas. Position the cuvette in the measurement chamber so that the graduation aligns with the white triangle on the housing marking. Always use the cuvette lid with a sealing ring to prevent contamination and ensure accurate readings.

What are the safety precautions for using the photometer?

Avoid direct sunlight when operating the device. Keep the cuvette and lid clean to prevent measurement errors. Do not touch the reagent tablets with your fingers. Store reagent solutions properly and dispose of them according to local regulations. Always follow the manufacturer's instructions for safe and accurate operation.

How to replace the batteries?

To replace the batteries, open the battery compartment cover, insert new AA batteries (LR03), and close the compartment. Ensure the sealing ring is properly seated to maintain the device's integrity. If the batteries are removed for more than one minute, the date and time settings will reset upon re-powering.

How to maintain the photometer for accurate results?

Regularly clean the transparent measurement chamber and check the light transmission areas for contamination. Use lint-free cloths and cotton swabs for cleaning. Avoid temperature fluctuations and ensure the device is used in a stable environment. Always calibrate the device before each measurement and follow the recommended maintenance schedule.