What is the principle of operation for the LF module?
The LF module operates using an electrodeless, inductive measurement principle. It measures the electrical conductivity of liquids based on their dissociation level, which is expressed in mS/cm. This allows for the determination of solution concentration.
How does temperature compensation work in the LF module?
The LF module includes a built-in temperature sensor to compensate for temperature effects on measurement results. The temperature compensation factor can be set between 0%/K and 5%/K in increments of 0.01%/K. Accurate measurement requires individual calibration of the temperature compensation factor (Tkα) for the solution being measured.
What are the electrical connection requirements for the LF module?
Only Ecolab Engineering GmbH sensors should be used. The standard setup includes 10 m of cable. For longer distances, a shielded special cable (Artikel-Nr. 418437041) and a special terminal box (Artikel-Nr. 288101) are required. The maximum cable length should not exceed 100 m.
How should the sensor be installed in a pipe?
The sensor should be installed in the flow direction, with the cross-hole aligned parallel to the pipe axis. The distance between the sensor and the device should not exceed 10 m. The installation position should ensure proper mixing in the measurement area.
What are the installation requirements for the tank wall?
The installation location must ensure sufficient mixing in the measurement area. The cross-hole of the sensor should be aligned vertically or up to a maximum angle of 45°. The distance between the sensor and the device should not exceed 10 m.
How is the sensor installed using a weld-on fitting?
A hole of 49 mm diameter is drilled in the tank wall. The weld-on fitting is properly welded to the tank wall. The sensor is then inserted into the fitting and secured with a cover nut. Proper material selection is essential to prevent corrosion at the weld joint.
How is the sensor installed using a PVDF cover nut?
A hole of 21 mm diameter is drilled in the tank wall. The sensor is inserted from the inside of the tank through the flat gasket into the hole. The sensor is then secured with a cover nut that includes a gasket. This method is suitable for straight tank walls.
How is the immersion pipe installed?
The immersion pipe is installed by drilling a 49 mm hole in the tank lid. The weld-on fitting is properly welded to the lid. The immersion pipe is then inserted into the fitting from the top and secured with a cover nut. The distance between the sensor and the controller should not exceed 4 m.
What are the maintenance requirements for the LF module?
Regular maintenance is required to ensure the proper functioning of the LF module. This includes checking the connections, cleaning the sensor, and verifying the calibration settings. The module should be inspected periodically for signs of wear or damage.
What should be done if the LF module is not functioning properly?
If the LF module is not functioning properly, the first step is to check the connections and ensure all parameters are correctly set. If the issue persists, the module should be inspected for damage or wear. If the device shows visible damage or no functionality, it should be taken out of service and secured against unintended operation.