Kamstrup MULTICAL 402 (114 pages)
Tags
Languages available:
No exact matches found. Please change your query
Need help?
Do you have a question about the product that is not answered in the manual?
- Receive updates on solutions
- Get answers to your question
Product description:
Frequently Asked Questions
What is the operating temperature range for the heat meter?
The heat meter operates within a temperature range of 2 °C to 160 °C for the computer unit, 10 °C to 150 °C for the temperature sensor pair, and 15 °C to 130 °C for the volume measuring section. The ambient temperature should be between 5 °C and 55 °C.
How do you start the radio function of the meter?
To start the radio function immediately, press both operating buttons simultaneously for 5 seconds. The confirmation of the radio function is indicated by a 3-second display showing 'CALL' on the screen.
What is the default start date for the meter?
The default start date for the meter is January 1st. The start date value is calculated at midnight between December 31st and January 1st. This date can only be changed using the manufacturer's hardware and software.
What are the installation requirements for the volume measuring section?
The volume measuring section must be installed in a dry, easily accessible location. Ensure that the flow direction arrow on the meter aligns with the actual flow direction. Avoid installation at low points and ensure proper support for the piping.
What is the minimum distance between the computer unit and electromagnetic interference sources?
The computer unit must be installed at least 50 cm away from any potential sources of electromagnetic interference to prevent measurement errors and ensure accurate readings.
What are the recommended installation positions for the volume measuring section?
The volume measuring section can be installed horizontally, vertically, or at an angle. It can be rotated up to 45° upwards or 90° downwards relative to the pipe axis. In damp environments, the meter should be installed at a 45° angle to the pipe axis.
What is the minimum flow rate for the meter?
The minimum flow rate for the meter is 0.6 m³/h. The meter is designed for use with water without additives, and it is not suitable for glycol-water mixtures.
How should the meter be cleaned?
The meter should only be cleaned from the outside using a soft, slightly damp cloth. Avoid using any harsh chemicals or abrasive materials that could damage the surface or affect its performance.
What are the safety precautions for installation?
Ensure the piping system is properly grounded and that lightning protection is provided through the house installation. Avoid welding, soldering, or drilling near the meter during installation. Always use appropriate safety equipment and follow local electrical codes.
What is the purpose of the temperature sensors?
The temperature sensors are used to measure the temperature difference between the supply and return lines, which is essential for accurate energy consumption calculations. They must be installed symmetrically and ensure that their ends reach at least the middle of the pipe cross-section.