OHAUS MB27 (64 pages)
Tags
Languages available:
No exact matches found. Please change your query
Need help?
Do you have a question about the product that is not answered in the manual?
- Receive updates on solutions
- Get answers to your question
Product description:
Frequently Asked Questions
How do I calibrate the moisture analyzer?
To calibrate the moisture analyzer, first remove any load from the pan support. Then, press and hold the Cal button in Weigh mode until 'CAL' appears. Place a 50g calibration mass on the pan support and follow the on-screen instructions to complete the calibration process.
What is the recommended sample weight for accurate results?
For accurate and repeatable results, the sample weight should generally be between 3g and 20g. Smaller samples may sacrifice some accuracy, while larger samples provide more consistent results but require longer test times.
How do I set the drying temperature and time?
To set the drying temperature and time, navigate to the Edit mode using the Temp or Time buttons. Use the Set buttons to adjust the values, and press Tare to confirm your selection. Ensure the settings are appropriate for the type of sample being tested.
What should I do if the display shows 'Close Cover'?
If the display shows 'Close Cover,' it means the heater cover is not properly closed. Ensure the cover is fully closed before starting the test to allow the analyzer to function correctly and avoid errors during the process.
How do I clean the moisture analyzer?
Before cleaning, disconnect the analyzer from the power supply and ensure it is fully cooled. Use a lint-free cloth slightly dampened with water or a mild cleaning agent to clean the housing and temperature sensor. Avoid using solvents or abrasive cleaners.
What should I do if the analyzer is not turning on?
If the analyzer is not turning on, check the power connections, fuse, and voltage. Ensure the power cable is properly connected to a grounded outlet and that the input voltage matches the local power supply specifications.
How do I prepare a sample for testing?
To prepare a sample, place the empty test pan on the pan support and press Tare to zero the pan weight. Add the sample, ensuring it is greater than 0.5g and evenly spread across the pan. Place the pan back on the support and proceed with the test.
What are the safety precautions for using the analyzer?
Always use the analyzer in a safe, well-ventilated area. Avoid operating it in hazardous or unstable environments. Never place flammable materials near the analyzer, and ensure the area has sufficient free space around the device for safety.
How do I change the displayed result unit?
To change the displayed result unit, press the %g button. This will toggle between grams, percent moisture, and percent solids, allowing you to view the results in the desired format.
What should I do if the analyzer shows poor accuracy?
If the analyzer shows poor accuracy, ensure it is placed in a stable environment with minimal vibrations and temperature fluctuations. Perform a calibration check and verify that the sample preparation and settings are appropriate for the type of sample being tested.