Product description:

The techem compact V is a compact measuring instrument designed for precise and reliable measurements. It features a user-friendly interface and is suitable for a wide range of applications in various industries. The device is built with high-quality materials to ensure durability and long-term performance. It supports multiple measurement functions and provides accurate results even in challenging environments. Its compact design makes it easy to handle and transport, making it ideal for on-site use.

Frequently Asked Questions

What are the installation requirements for the heat meter?

The heat meter must be protected against magnetite and dirt for its entire lifespan. Shut-off valves must be installed in front of the dirt catcher and behind the meter. It should only be installed in the section for which it is designed, with the outlet variant in the low-temperature section and the inlet variant in the high-temperature section.

How to install the measuring capsule?

Thoroughly rinse the pipeline, close the shut-off valves in the inlet and outlet, unscrew the temporary cover or old meter anti-clockwise, clean the sealing surfaces of the connector and the O-ring of the measuring capsule, and lightly grease the sealing surfaces with tap grease.

How to install the temperature sensors?

Remove pressure from the sensor installation location, unscrew the closing screw from the special ball valve, place the O-ring on the assembly pin, insert the O-ring into the bore hole of the closing screw, position the O-ring with the assembly pin, place the sleeve end over the temperature sensor, and secure the temperature sensor with the plastic screw.

What are the display levels for the heat meter?

The heat meter has two display levels: the readout level and the service level. A long button press switches between the levels, while a short press jumps to the next display. The readout level shows cumulative energy and current flow, while the service level provides detailed information like temperature readings and firmware details.

What does the error code F-1 indicate?

Error code F-1 indicates a faulty temperature sensor. The device must be replaced and returned for further inspection and repair.

What does the error code F-3 indicate?

Error code F-3 indicates that the return flow sensor is registering a higher temperature than the inlet sensor. Check whether the heat meter and sensors are installed in the correct sections to resolve the issue.

What does the error code F-4 indicate?

Error code F-4 indicates a faulty flow sensor. The device must be replaced and returned for further inspection and repair.

What does the error code F-5 indicate?

Error code F-5 indicates that the heat meter is measuring correctly. To save power, the optical interface is temporarily disabled, and no data transmission occurs.

What does the error code C-1 indicate?

Error code C-1 indicates a permanent fault in the meter. The device must be replaced, and the readout values cannot be used due to the fault.

How to perform a functional check of the heat meter?

Open the shut-off valves in the inlet and outlet, check the leak tightness of the connector, and press the push-button on the meter to activate the display. Ensure all components are functioning correctly and the meter is properly sealed.

Send answer