VWR DO220 (156 pages)
Tags
Languages available:
No exact matches found. Please change your query
Need help?
Do you have a question about the product that is not answered in the manual?
- Receive updates on solutions
- Get answers to your question
Product description:
Frequently Asked Questions
How do I install the battery in the VWR DO220?
To install or replace the battery, remove the two battery cover screws and the battery cover. Insert a new 9V battery, then replace the battery cover and o-ring, ensuring the o-ring is correctly aligned to prevent a poor seal. Secure the screws and check the LCD for a 'LOW BAT' indication, which means the battery will last approximately one hour.
What are the temperature specifications for the VWR DO220?
The VWR DO220 measures temperature in the range of 0 to 50°C with a resolution of 0.1°C and an accuracy of ± 0.3°C. It conforms to EMC 2004/108/EC and RoHS 2011/65/EU directives, as well as several IEC harmonized standards.
How do I store the VWR DO220 probe for long-term use?
For long-term storage, keep the probe in the gray calibration/storage sleeve with a moistened, clean sponge. This prevents the sensor cap from drying out. If the sensor cap dries, rehydrate it using the instructions in the Maintenance and Troubleshooting section.
What is the battery life of the VWR DO220 in manual sampling mode?
The VWR DO220 uses a 9V alkaline battery with a battery life of up to 100 hours in manual sampling mode. Replace the battery when the LCD displays 'LOW BAT', indicating the battery will last approximately one hour.
How do I prepare the probe for the first time?
To prepare the probe for the first time, remove the probe guard by turning it counterclockwise, then take off the red storage cap containing a moist sponge. Reinstall the probe guard by sliding it over the sensor and threading it clockwise. Place the probe in the gray calibration/storage sleeve with a moist sponge for storage and calibration.
What are the dissolved oxygen measurement ranges for the VWR DO220?
The VWR DO220 measures dissolved oxygen in the range of 0.0 to 200% air saturation or 0.00 to 20.0 mg/L. It has a resolution of 1% or 0.1% (user selectable) and an accuracy of ± 1.5% of reading or ± 1.5 mg/L, whichever is greater.
How do I calibrate the VWR DO220?
Calibration is required for accurate measurements. Follow the instructions in the Calibration section of the manual, which includes steps for preparing the sensor cap, entering calibration coefficients, and rehydrating the sensor cap if necessary.
What should I do if the VWR DO220 gives unstable readings?
Unstable readings may indicate a damaged or dirty sensor cap. Clean or replace the sensor cap using the instructions in the Maintenance and Troubleshooting section. If the issue persists, check for scratches on the paint or dye layer and replace the sensor cap if needed.
How often should I replace the sensor cap on the VWR DO220?
The average sensor cap replacement interval is 12 to 18 months, although they may last longer. Replace the sensor cap if you notice erratic readings, large scratches on the paint or dye layer, or if the sensor cap dries out. Follow the instructions in the Maintenance and Troubleshooting section for cleaning or replacement.
What are the key advantages of the VWR DO220?
The VWR DO220 offers several advantages, including the elimination of sensor flow dependence (stirring), sensor warm-up time, and frequent membrane/electrolyte changes. It provides greater stability, less frequent calibrations, and accurate measurements of temperature and dissolved oxygen in both % and mg/L units.