Product description:
Frequently Asked Questions
What is the power supply voltage for the Mantis Compact?
The power supply voltage for the Mantis Compact is 100-240VAC 50-60HZ with a maximum current of 1.0A. The stand input voltage is 9VDC 1.7A.
What is the operating temperature range for the Mantis Elite?
The operating temperature range for the Mantis Elite is between 0°C and +40°C. The storage temperature range is 0°C to +50°C for up to 3 months without adverse effects.
How do you adjust the eye spacing on the Mantis Compact?
To adjust the eye spacing on the Mantis Compact, use the eye spacing control to obtain a comfortable stereo view. This adjustment is crucial for viewing comfort and should be customized for each user.
What is the working distance for the x4 objective lens on the Mantis Compact?
The working distance for the x4 objective lens on the Mantis Compact is 96mm. This allows for close-up viewing while maintaining a clear field of view.
How do you change the objective lens on the Mantis Elite?
To change the objective lens on the Mantis Elite, loosen the lens securing screw, remove the lens, and store it in the supplied holder. Fit the chosen lens into the objective seat and tighten the securing screw. Alternatively, rotate the objective turret to switch lenses.
What is the maximum relative humidity for the Mantis Compact?
The maximum relative humidity for the Mantis Compact is 80% for temperatures up to 31°C. It decreases linearly to 50% relative humidity at 40°C. Ensure the equipment is used in environments within these limits.
How do you perform desiccant replacement on the Mantis Compact?
To replace the desiccant on the Mantis Compact, disconnect the power supply, remove the screw securing the desiccant cover, lift off the cover, insert the new desiccant, and replace the cover. Ensure the desiccant is properly seated to maintain optimal moisture control.
What is the warranty period for the Mantis Elite?
The Mantis Elite is warranted to be free from defects in material and workmanship for a period of one year from the date of invoice to the original purchaser. If defective, it will be repaired or replaced at Vision Engineering's option, with certain exceptions for expendable components.
How do you adjust the head counterbalance on the Mantis Compact?
To adjust the head counterbalance on the Mantis Compact, use the counterbalance screw. Turn it clockwise for lighter lenses or counter-clockwise for heavier ones until the viewing head does not drift up or down when unsupported.
What is the luminous intensity of the Mantis Compact?
The luminous intensity of the Mantis Compact is 9,400 LUX when measured at the subject plane with colour correction filters. This ensures high-quality illumination for detailed viewing.