Frequently Asked Questions
What materials are used in the wetted parts of the sensor?
The wetted materials vary depending on the sensor model. For example, the 8310 model uses a black PSU body, stainless steel 316L internal and external electrodes, PSU insulator, and a glass polyester/IP65 connector. The 8315 model uses a stainless steel 316L body, stainless steel 316L internal and external electrodes, PES insulator, Viton o-ring, and glass polyester/IP65 connector.
What is the measuring range of the sensor?
The measuring range depends on the cell constant. For a cell constant of 0.01, the range is 0.01–200 µS/cm. For a cell constant of 0.1, the range is 0.1 µS–2 mS/cm. For a cell constant of 1, the range is 1 µS–20 mS/cm.
What is the response time of the sensor?
The response time of the sensor is 0.5 seconds, ensuring quick and accurate readings during dynamic process conditions.
What is the repeatability and precision of the sensor?
The repeatability is ±0.1 µS/cm, and the precision is ±0.5% of the reading for values between 20–200,000 µS/cm. These specifications ensure reliable and consistent measurements across a wide range of conductivity levels.
What is the maximum cable length for the sensor?
The maximum cable length for the sensor is 91 meters (299 feet), allowing for flexible installation in various environments while maintaining signal integrity.
What is the temperature element used in the sensor?
The temperature element used in the sensor is a PT100, which provides accurate temperature compensation for conductivity measurements. Some models may also support a PT1000 element for enhanced precision.
How do you configure the sensor for different measurement parameters?
To configure the sensor, use the Configure menu to select the desired parameter, such as conductivity, TDS, salinity, or resistivity. You can also set the display format, units, and cell constant to match your application requirements.
How do you set the temperature compensation for the sensor?
Temperature compensation can be set to linear, ammonia, temp table, or natural water/pure water modes. For custom applications, a user-defined linear compensation can be entered with a specified slope and temperature range to improve measurement accuracy.
What is the purpose of the filter setting in the sensor?
The filter setting adjusts the time constant to improve signal stability. A longer time constant averages the signal over a specified period (up to 60 seconds), reducing noise and ensuring more stable readings during fluctuating process conditions.
What are the safety precautions when installing the sensor?
When installing the sensor, ensure the power is disconnected to avoid electrical hazards. Use industry-approved hardware and follow safety procedures for high-pressure and high-temperature environments. Always refer to the manual for specific installation instructions and safety alerts.