What is the maximum temperature for drying the instrument?
The instrument must never be heated above 138°C during the drying process, whether using automated or manual drying methods.
How should the instrument be dried manually?
Dry the outside of the instrument with a lint-free towel. For the lumens, use filtered compressed air (max. 3 bars).
What is the recommended cleaning procedure for the instrument?
After use, wipe the instrument to remove gross soiling. Rinse the water line with distilled or deionized water. For thorough cleaning, use a washer/disinfector with a Vario TD program, which includes pre-washing, washing, neutralizing, and intermediate rinsing steps.
What should be done if visible stains remain after cleaning?
If stains are still visible after cleaning or disinfection, the entire procedure must be repeated. If the product shows visible damage, chip/flake loss, corrosion, or deformation, it must be disposed of and not used further.
What is the recommended disinfection method for the instrument?
Disinfection should be performed using an aldehyde-free solution compatible with the product. For automated disinfection, use a washer/disinfector with a thermal cycle of at least 93°C for 3 minutes, following national A0-value requirements.
How should the instrument be prepared for sterilization?
Before sterilization, place the instrument in a suitable sterilization container or packaging that complies with EN ISO 11607 or EN 868 standards. Ensure the disinfection step is completed no later than 2 hours after cleaning.
What is the recommended method for pre-cleaning the instrument?
Rinse the surface of the instrument with water or a cleaning agent. Rinse all lumina (e.g., irrigation and aspiration connections) at least three times in the normal flow direction using a disposable syringe filled with distilled/deionized water. Alternatively, use an aldehyde-free cleaning and disinfection solution, followed by thorough rinsing with water.
What is the recommended storage condition for the instrument?
Store the instrument in a clean, dry environment to prevent contamination and damage. Ensure it is properly packaged and protected from environmental factors during storage.
What should be done if the instrument is damaged?
If the instrument is damaged, it should be reprocessed before sending it back to the manufacturer for repair. Do not use the device if it shows visible damage, chip/flake loss, corrosion, or deformation.
What is the recommended method for functional testing and maintenance?
After cleaning and disinfection, perform a functional test to ensure the instrument operates correctly. Regular maintenance should be carried out according to the manufacturer's guidelines to ensure optimal performance and longevity.