Product description:

The Aim-TTI LD300 is a high-precision power supply designed for laboratory and industrial applications. It provides stable and accurate voltage and current outputs, making it suitable for testing and calibrating electronic components and devices. The unit features a digital display for easy monitoring of output parameters and includes multiple protection functions to ensure safe operation. Its compact design and robust construction allow for reliable performance in various environments.

Frequently Asked Questions

What is the maximum current rating of the LD300?

The LD300 DC Electronic Load has a maximum current rating of 80 Amps at the rear panel terminals and 30 Amps at the front panel terminals.

What are the operating voltage limits for the LD300?

The LD300 operates over a voltage range of 0 to 80 Volts. Surge suppressors activate at 120V, and the maximum non-repetitive surge energy is 80 Joules.

What is the power dissipation capability of the LD300?

The LD300 has a power dissipation capability of up to 300 Watts at 28°C, with a derating to 300 Watts at 40°C.

What are the operating modes of the LD300?

The LD300 offers five operating modes: constant current, constant voltage, constant power, constant resistance, and constant conductance. Each mode is set using 10-turn controls on the front panel.

How does the LD300 handle transient control?

The LD300 includes a transient generator that can switch the load between two levels set by 10-turn controls. The frequency and duty cycle of the transients are adjustable, and transitions have a linear slewing characteristic.

What is the accuracy of the current measurement in constant current mode?

In constant current mode, the current accuracy is ± (0·2% + 30 mA), with a regulation of < 30 mA for a 90% load power change.

What is the accuracy of the voltage measurement in constant voltage mode?

In constant voltage mode, the voltage accuracy is ± (0·2% + 2 digits), with a regulation of < 30 mV for a 90% load power change.

What is the maximum power range of the LD300?

The LD300 can handle a power range of 0 to 320 Watts, with a setting accuracy of ± (0·5% + 2 W) and a regulation of < 2% over a 5 V to 75 V source voltage change.

What is the dropout voltage feature of the LD300?

The LD300 has a low voltage dropout facility that protects sources like batteries by reducing the load current when the source voltage falls below the dropout threshold. The setting accuracy is ± 1% + 2 digits.

What are the environmental operating conditions for the LD300?

The LD300 operates within a temperature range of +5°C to +40°C and a relative humidity range of 20% to 80% (non-condensing). It is designed for indoor use at altitudes up to 2000 meters.