Product description:

The WIKA CPP1000-M is a high-precision pressure pump designed for use in calibration and testing applications. It features a robust construction and precise pressure control, allowing for accurate and repeatable pressure generation. The pump is suitable for a wide range of media, including gases and liquids, and is equipped with a pressure display for real-time monitoring. Its compact design and ease of operation make it a reliable choice for laboratories and industrial environments.

Frequently Asked Questions

What is the maximum pressure the CPP1000-M/-L can handle?

The maximum permissible pressure for the CPP1000-M/-L is 1000 bar. Higher pressures may damage the pump, so ensure all connected instruments and tubes are rated for this pressure.

How do I set up the hand spindle pump?

Place the hand spindle pump on a solid surface. Secure it with appropriate screws if needed. Ensure the star handle is properly seated and the spring-loaded thrust pad engages into the bushing.

What type of operating liquid should I use?

Use mineral oil-based hydraulic fluid or clean water free of calcium-carbonate/scale. Distilled water and water-based fluids are not suitable. Other pressure transmitting media can be requested.

How do I connect pressure measuring instruments?

Connect the reference and test instruments to the two pressure connections. Ensure the O-ring seals are in place and tighten with hand pressure. Use threaded adapters if the instruments have different connection threads.

How do I vent the system before calibration?

Open the shut-off valve and turn the star handle anticlockwise until fully out. Close the valve, then slowly turn the star handle clockwise until pressure reaches approximately 50 to 100 bar. Repeat if necessary to remove trapped air.

How do I adjust the test pressure?

To increase test pressure, turn the star handle clockwise. To decrease it, turn anticlockwise. Use the fine adjustment valve for precise control of the initial pressure setting.

What should I do if the pump does not build pressure?

If no pressure is generated after fully winding the spindle clockwise, the connected test volume may be too large. Pre-fill the test specimen with operating liquid or use the shut-off valve for test connection as an accessory.

Can I use the pump in different environments?

The hand spindle pump should be handled with care, avoiding humidity, impacts, and extreme temperatures. Ensure the device temperature adjusts to ambient conditions when moving from cold to warm environments.

What safety precautions should I follow?

Only trained and authorized personnel should operate the pump. Follow all safety instructions in the manual. Do not attempt to repair or alter the device yourself. Use original seals and avoid applying force to control elements.

How do I maintain the pump?

Regularly check the O-ring seals for wear and replace if necessary. Ensure the operating liquid is clean and replace it as needed. Follow the maintenance guidelines in the manual to ensure reliable operation and safety.