Product description:
Frequently Asked Questions
What are the differences between M1196A and M1196T sensors?
The M1196A and M1196T sensors have identical physical appearances except for cable length and connector type. The M1196A uses an 8-pin connector with a 3-meter cable, while the M1196T uses a 9-pin D-sub connector with a 0.9-meter cable. Both are intended for adult patients weighing over 40kg.
Which instruments are compatible with M1196A and M1196T sensors?
The M1196A is compatible only with validated Philips/Agilent/HP instruments. The M1196T is compatible with validated Philips/Agilent/HP instruments and compatible OEM instruments. Refer to the Sensor Compatibility Sheet for a full list of compatible devices.
How do I apply the clip sensor to a patient?
To apply the sensor, open the padded finger cavity by pressing the spring-tabs, guide the sensor over the fingertip, and secure it by releasing the tabs. Ensure the sensor cable rests on top of the patient’s hand. Adjust the sensor position as needed for comfort and proper fit.
What should I do if the sensor is too tight?
If the sensor is too tight, it may cause venous congestion and tissue ischemia. Ensure the sensor is not overly tight by checking the application site for signs of pressure marks or reduced circulation. Adjust the sensor position or size if necessary.
How often should I change the sensor application site?
Move the sensor application site every 4 hours, or more frequently if circulation or skin integrity is compromised. Prolonged use on a single site may lead to skin irritation or ulceration.
What cleaning solutions are recommended for the sensors?
Use only validated cleaning solutions such as 70% isopropanol or isopropanol wipes. Avoid immersion in any liquid, and never use bleach or sodium hypochlorite-based products. Always clean and disinfect the sensor before reusing it on a different patient.
What is the accuracy of the M1196A and M1196T sensors?
The M1196A and M1196T sensors have an accuracy of ±3% SpO2 in the range from 70% to 100% SpO2. For more detailed specifications, refer to the Sensor Compatibility Sheet included with the sensor.
Can I use the sensor during MRI scans?
No, do not use the sensor during MRI scanning. This may cause burns or inaccurate measurements. Ensure the sensor is removed before any MRI procedures to avoid potential harm to the patient.
What should I do if the sensor shows signs of damage?
If the sensor shows any signs of damage or alteration, it should not be used for further patient monitoring. Dispose of the sensor according to proper disposal procedures to prevent any risk to patient safety.
What are the light emitting diode wavelength ranges for these sensors?
The light emitting diodes used in these sensors operate within a wavelength range of 600nm to 1000nm, with an optical output power of less than 15mW. This range is useful for clinicians involved in photodynamic therapy.