Product description:
Frequently Asked Questions
How do I prepare the test cartridge?
To prepare the test cartridge, press the power button to turn on the analyser and wait for the initialisation. Remove the test cartridge from its foil wrapper, check that the silica gel pack is present, intact, and yellow. If the silica gel is missing, open, or green, discard the cartridge and start again with a new one.
How do I collect the oral fluid sample?
Ask the donor to unwrap a new collection device and hold it by the plastic stem. Place the device in their mouth and actively swab around the gums, tongue, and inside the cheek until the sample presence indicator starts to turn blue.
How do I run the test?
Insert the collection device into the test cartridge in the analyser and push it all the way in until it stops. The analyser will then test the sample, displaying the test time on the screen. Keep the analyser horizontal and still throughout the process.
What should I do after the test completes?
The results will be displayed on the analyser screen. If enabled, a donor questionnaire will start after pressing 'OK'. You can print the results if required, but ensure the printer is connected and turned on before printing. Press 'NO' to skip this step.
What are the storage conditions for the test kit?
Store all contents at 15°C to 25°C. Do not freeze. The reagents are stable until the expiry date on the packaging, and the test cartridge shelf life is indicated in the barcode, which the analyser will read upon insertion.
What are the cutoff levels for drug detection?
The cutoff levels for drug detection are as follows: Amphetamine (50 ng/mL), Benzodiazepines (20 ng/mL), Cannabis (25 ng/mL), Cocaine metabolite (30 ng/mL), Methadone (15 ng/mL), Methamphetamine (50 ng/mL), and Opiates (40 ng/mL).
How does the analyser interpret test results?
The analyser interprets the line intensity on the test strip, comparing it with a predetermined threshold derived from concentration curves. It reports qualitative results as negative, positive, or lab test required. Results can be displayed as positive/negative or lab test required/negative, depending on the analyser settings.
What are the limitations of the test?
This test is for professional use only and provides preliminary results. Positive results should be confirmed with a more specific method like GC-MS or LC-MS. The test does not distinguish between drugs of abuse and certain medications, and results may be affected by technical errors or interfering substances.
What is the sensitivity and specificity of the test?
The SoToxa™ Test Kit has a sensitivity of ≥90%, specificity of ≥99%, and accuracy of ≥95%. These performance metrics were determined in a controlled study using drug-free samples and samples spiked at 100% of the respective cutoff levels.
What substances may cause false positives?
Several substances, including codeine, heroin, morphine, and others, may cause false positives at high concentrations. The test may also react to certain medications, and cross-reactivity data is provided for various compounds. Always consult the manual for a complete list of potential interfering substances.