Tyco KENDALL GENIUS 2 (242 pages)
Tags
Languages available:
No exact matches found. Please change your query
Need help?
Do you have a question about the product that is not answered in the manual?
- Receive updates on solutions
- Get answers to your question
Product description:
Frequently Asked Questions
How long should the GENIUS 2 Checker/Calibrator warm up before use?
The GENIUS 2 Checker/Calibrator should warm up for at least 15 minutes after power is applied. This allows the infrared blackbody targets to reach the correct operating temperatures.
What is the purpose of the GENIUS 2 Checker/Calibrator?
The GENIUS 2 Checker/Calibrator is used to check the accuracy of the GENIUS 2 Tympanic Thermometer and automatically recalibrate it if necessary. It also generates a test report for each thermometer tested and saves it to a USB flash drive.
What are the two infrared calibration targets used in the GENIUS 2 Checker/Calibrator?
The GENIUS 2 Checker/Calibrator uses two infrared calibration targets: a low temperature target (32.2° C / 90.0° F) and a high temperature target (40.6° C / 105.0° F). These targets simulate the heat transfer of the human ear.
What should be done if the GENIUS 2 Thermometer lens is dirty?
If the thermometer lens is dirty, it should be cleaned thoroughly before use. This ensures accurate calibration checks. Refer to the cleaning instructions in the GENIUS 2 Thermometer operating manual for proper cleaning procedures.
What is the recommended procedure for inserting the probe tip into the target?
The probe tip should be inserted into the target within 60 seconds of pressing the encoder knob. Ensure the probe tip is fully inserted into the target well and avoid pointing it at any object, including the LCD display, to prevent inaccurate readings.
What happens if the GENIUS 2 Thermometer is found to be outside the specified accuracy limits?
If the thermometer is found to be outside the specified accuracy limits, the GENIUS 2 Checker/Calibrator will initiate a recalibration sequence. The user must repeat Steps 3 through 6 three more times to gather recalibration data. If recalibration is successful, a results screen will indicate the thermometer passed calibration.
What should be done if the GENIUS 2 Thermometer fails to communicate with the Checker/Calibrator?
If communication fails, check both ends of the thermometer probe cable, especially the end inside the battery compartment. Ensure the pins are not bent or damaged. If the problem persists, try a different thermometer or return the unit for factory servicing.
What is the purpose of the USB flash drive in the GENIUS 2 Checker/Calibrator?
The USB flash drive is used to store test reports generated by the GENIUS 2 Checker/Calibrator. The report includes the thermometer serial number, date, time, user name, and other relevant information. The file is saved in a simple ASCII format and is only available in English.
What should be done if the GENIUS 2 Checker/Calibrator displays an 'Error 1: Internal Error'?
An 'Error 1: Internal Error' indicates a problem with the Checker/Calibrator itself. If the error persists after cycling the power, the unit should be returned to the factory for servicing.
What is the recommended procedure for cleaning the GENIUS 2 Thermometer?
Before using the GENIUS 2 Thermometer, ensure the lens is clean. Use the cleaning instructions provided in the GENIUS 2 Thermometer operating manual. Avoid using any contaminants or substances that could affect the accuracy of the thermometer.