Product description:
Frequently Asked Questions
How does the voltage and socket test work?
The voltage and socket test measures the voltage between phase and neutral and checks the socket wiring according to local requirements. If the wiring is correct and the voltage contact is less than 50V, the display shows a green bar. If there is an error, the bar turns red and an alarm sounds.
What is the purpose of the RCD test?
The RCD test checks the residual current device according to EC/EN 61557-6 standards. It injects a test current of 30 mA through the ground for type A, AC, and F. The test includes tripping time and contact voltage measurements to ensure the RCD functions correctly.
How to perform the RCD test in AUTO mode?
In AUTO mode, the instrument performs a sequence of tests: x1/2 at 0°, x1/2 at 180°, x1 at 0°, x1 at 180°, x5 at 0°, and x5 at 180°. The display shows the tripping time in milliseconds. A green bar indicates the test meets local regulations, while a red bar indicates failure.
What is the function of the LOOP test?
The LOOP test measures impedance between phase and ground without tripping the RCD. It follows IEC/EN61557-3 standards and can be used for TT systems to measure earth resistance and contact voltage, or for TN systems to measure impedance and prospective short circuit current.
How to select the TT or TN mode for the LOOP test?
To select TT or TN mode for the LOOP test, plug the instrument into the socket and press the FUNC button to select LOOP. Use the arrow button to choose TT or TN, and the button to select TEST STD or LOW. Press the RED button for 2 seconds to start the measurement.
What is the maximum voltage input for the instrument?
The maximum voltage input for the instrument is 265 V. Connecting the instrument to a higher voltage can cause permanent damage and electrical shock. Always ensure the voltage is within the specified limit to avoid safety hazards.
How does the supercap technology work?
The supercap technology allows the instrument to retain power even when disconnected from the socket. A fully charged supercap enables the instrument to operate for 45 seconds after being disconnected, allowing the user to read results after an RCD trip or when the display is difficult to read.
What are the display colors indicating?
Green indicates that the measured value is correct according to local regulations, while red indicates that the test was not passed. During the socket test, an incorrect value also triggers a sound alarm to alert the user.
What is the purpose of the harmonics test?
The harmonics test measures voltage harmonics up to the 49th harmonic and calculates the total harmonic distortion (THD%). This helps assess the quality of the electrical supply and identify potential issues caused by harmonic distortion.
How to access the memory and reports?
The instrument has internal memory to store test results. The subfeatures selection button allows users to access stored data or analyze harmonics up to the 49th harmonic. The memory can be reviewed for historical test data and reports.