Product description:
Frequently Asked Questions
How to replace the battery?
To replace the battery, open the battery compartment by folding the cover away from the handle. Replace the old battery with a new one of the same type and fold the cover back into place. Ensure the battery is fully charged and avoid leaving batteries loose where they can be accessed by children.
What is the temperature measurement range?
The infrared thermometer can measure temperatures ranging from -35 °C to +800 °C (-21 °F to +1,472 °F). Accuracy varies depending on the temperature range, with ±2.5 °C (±4.5 °F) for temperatures between -35 °C and +20 °C, and ±1% of the measured value for higher ranges.
How to adjust the emissivity setting?
To adjust the emissivity setting, navigate to the appropriate function mode by pressing the mode button. Use the up and down buttons to set the desired emissivity value between 0.10 and 1.0. This setting remains stored even when the device is turned off.
What are the technical specifications of the device?
The device features a 1999-count display with backlighting, a thermopile detector, and a temperature range of -35 °C to +800 °C. It has a resolution of 0.1 °C (0.1 °F), a laser class 2 (II) indicator, and an optical resolution of 12:1. The minimum spot size is 12.5 mm, and it automatically turns off after about 10 seconds.
What materials have the highest emissivity?
Materials with the highest emissivity include non-metallic colors (0.90–0.95), non-alkaline colors (0.90–0.95), and black enamel paint (0.95). These materials emit a high percentage of thermal radiation and are ideal for accurate temperature measurements.
How to use the continuous measurement function?
To activate continuous measurement, press the mode button repeatedly until the corresponding function symbol blinks on the display. Use the up or down button to activate the continuous measurement mode. This function continuously measures temperature without requiring the measurement button to be pressed.
What should I do if the display shows '– – –'?
If the display shows '– – –', it means the target temperature is outside the measurable range. Ensure the target is within the device's temperature range (-35 °C to +800 °C). If the issue persists, check the battery level and replace it if necessary.
How to clean the device?
To clean the device, gently blow away loose dust from the IR lens. Use a soft brush to remove remaining dirt. Wipe the surface with a slightly damp cloth, using only water for dampening. Avoid using chemicals or cleaning agents.
What is the accuracy of the device in different temperature ranges?
The device has an accuracy of ±2.5 °C (±4.5 °F) for temperatures between -35 °C and +20 °C. For temperatures between +21 °C and +200 °C, the accuracy is ±1% of the measured value. For higher ranges, it is ±1 °C (±1.8 °F) or ±1.5% of the measured value, depending on the specific range.
How to set up the alarm function?
To set up the alarm function, navigate to the alarm mode by pressing the mode button. Use the up and down buttons to set the desired upper and lower alarm limits. The alarm settings are saved even when the device is turned off. The device will emit an acoustic alarm when the temperature exceeds the set limits.