Vaisala RFL100 (172 pages)
Tags
Languages available:
No exact matches found. Please change your query
Need help?
Do you have a question about the product that is not answered in the manual?
- Receive updates on solutions
- Get answers to your question
Product description:
Frequently Asked Questions
How do I set up RFL100 with one humidity or temperature probe?
To set up RFL100 with one humidity or temperature probe, first disconnect any cable from the service port. Open the battery cover, ensure the power switch is off, and connect the probe to the data logger. Align the orientation mark with the connector line, push the probe in without rotating, and close the battery cover. Verify the display shows the measurement reading instead of dashes or error codes.
What is required to connect two temperature probes to RFL100?
To connect two temperature probes, you need a probe splitter accessory (M8), two temperature probes (HMP110T, HMP115T, or TMP115), and optionally extension cables. Ensure the RFL100 firmware is at least version 1.2.0 and the viewLinc system is updated to firmware version 3.0 and software version 5.0.2. Follow the labeling steps to identify each probe's channel.
How do I connect a CO2 probe to RFL100?
To connect a CO2 probe, use the probe splitter M8/M12 and a GMP251 CO2 probe. Ensure the RFL100 firmware is at least version 1.4.0 and the device was manufactured after July 2021. Connect the probe to the M12 connector, use a micro-USB power supply, and ensure the data logger has lithium batteries for backup power. Check the display for error codes after setup.
What is the purpose of the probe splitter accessory?
The probe splitter accessory allows connecting multiple probes to the RFL100 data logger. It enables the use of two temperature probes or a CO2 probe with the data logger. The splitter must be aligned correctly with the probe connector, and it is essential for proper channel assignment and data collection.
What firmware versions are required for RFL100?
RFL100 requires firmware version 1.2.0 or higher for two temperature probes, and version 1.4.0 or higher for a CO2 probe. Additionally, the viewLinc system must have firmware version 3.0 for the AP10 access points and software version 5.0.2 for the viewLinc system to ensure compatibility and proper operation.
How do I ensure the RFL100 is properly connected to the network?
When RFL100 is first powered on, it scans for VaiNet access points in installation mode. It connects to the access point with the best signal strength and waits to be accepted by the viewLinc Enterprise Server administrator. Ensure the access points are installed and in installation mode before starting the setup process.
What should I do if the display shows dashes instead of measurements?
If the display shows dashes instead of measurements, check that the probe is properly connected. Ensure the probe or cable is not disconnected accidentally, and verify that the battery indicator shows full batteries. If the issue persists, consult the error codes section for further troubleshooting.
What is the role of the battery cover latch on RFL100?
The battery cover latch secures the battery cover in place after installation. To close the cover, push the latch down until you hear a click. If the cover does not close easily, push the probe or extension cable in slightly to ensure proper alignment and then attempt to close the cover again.
What are the power requirements for operating RFL100 with a CO2 probe?
RFL100 requires an external power supply when using a CO2 probe. Lithium batteries (type FR6) are recommended as a backup power source. If the external power fails, the lithium batteries will provide limited backup time, typically at least 12 hours. The data logger will display error code Err 106 if it operates in CO2 mode without a power supply.
How do I label the temperature probes for channel identification?
After connecting the first temperature probe, attach a label with number 1 to it. For the second probe, attach a label with number 2. If using TMP115 probes, use the included labels and connect them to the thin cable between the probe body and the sensor tip. This helps identify which probe is assigned to which channel for easier reference.