Product description:

The Lovibond CheckitDirect is a portable measuring instrument designed for the rapid and accurate determination of color and turbidity in liquids. It features a digital display, a compact design, and user-friendly operation. The device is suitable for use in various industries such as food and beverage, pharmaceuticals, and environmental monitoring. It provides reliable results with minimal sample preparation and is equipped with a built-in light source for consistent measurements.

Frequently Asked Questions

How to clean the cuvette after each analysis?

After each analysis, the cuvette, lid, and stirrer must be thoroughly cleaned to prevent carryover errors. Even small residues of reagents can lead to incorrect measurements.

What should the cuvette walls be like before analysis?

The outer walls of the cuvette must be clean and dry before analysis. Fingerprints or water droplets on the light transmission areas can cause measurement errors.

Why should the same cuvette be used for zero adjustment and testing?

The cuvettes may have slight tolerances between each other, so the same cuvette should be used for zero adjustment and testing to ensure consistent results.

How should the cuvette be positioned during zero adjustment and testing?

The cuvette must be placed in the measurement chamber so that the graduation aligns with the white triangle on the housing marking. This ensures accurate alignment for measurements.

What is the correct way to fill the cuvette?

The cuvette should be filled with the water sample up to the 10-ml mark. It must then be sealed with the lid and positioned correctly in the measurement chamber.

What is the procedure for zero adjustment and testing?

To perform zero adjustment and testing, the cuvette must be sealed with the lid, and the reagents must be added. The cuvette should be mixed thoroughly until the tablet dissolves, then placed back into the measurement chamber.

How to measure copper using the tablet method?

For free copper measurement, add a COPPER No. 1 tablet directly from the foil to the 10-ml sample, mix thoroughly, and then perform the zero adjustment and test. The result will be displayed in mg/l free copper.

How to measure total copper using the tablet method?

For total copper measurement, add a COPPER No. 2 tablet directly from the foil to the 10-ml sample, mix thoroughly, and then perform the zero adjustment and test. The result will be displayed in mg/l total copper.

What is the measurement tolerance for copper?

The measurement tolerance for copper is ±0.05 mg/l for concentrations between 0.05–1 mg/l and ±0.10 mg/l for concentrations between 1–5 mg/l.

How to measure pH using the tablet method?

For pH measurement, add a PHENOL RED PHOTOMETER tablet directly from the foil to the 10-ml sample, mix thoroughly, and then perform the zero adjustment and test. The result will be displayed as a pH value.