Product description:

The Lovibond TB 210 IR is a measuring instrument designed for the precise determination of color in liquids. It utilizes infrared technology to ensure accurate and reliable measurements, making it suitable for use in laboratories, quality control, and research environments. The device features a user-friendly interface and is built for durability, allowing for consistent performance over time. It is commonly used in industries such as food and beverage, pharmaceuticals, and chemical manufacturing, where color consistency is a critical quality parameter.

Frequently Asked Questions

How to dispose of electronic devices properly?

Tintometer GmbH provides a professional and environmentally responsible disposal service for electrical instruments purchased after August 13, 2005. This service is free of charge, excluding transportation costs. Send your used Tintometer instruments to your supplier for disposal.

What are the rules for disposing of batteries and accumulators?

According to the European Directive 2006/66/EC, used batteries and accumulators must not be disposed of with regular household waste. They should be returned to public collection points, shops selling batteries, or the supplier who provided them. This is a legal obligation for end users.

What safety precautions should be taken when using the turbidimeter?

Turbidimeter standards and T-CAL standards are for chemical analysis only and must not be accessible to children. Some standards contain substances that may be harmful to the environment. Always refer to safety data sheets for ingredients and dispose of standards properly. Ensure the device is used in an electromagnetically controlled environment.

How to prepare and use turbidity standards for regular use?

For regularly used standards, shake the standard vigorously for 2-3 minutes, let it rest for 5 minutes, and swirl the cuvette 5-10 times. Place the cuvette in the measurement chamber and wait for 1 minute before starting the measurement.

How to prepare and use turbidity standards for irregular use?

Newly delivered standards must rest for at least 24 hours. Do not shake or swirl <0.1 NTU standards. If the standard was shaken, allow several hours for air bubbles to dissipate. If the cuvette was accidentally swirled, let the standard rest for at least 15 minutes before use.

How to perform a routine measurement with the turbidimeter?

Turn the device on, clean the cuvette with the sample, fill it with the sample, and ensure it is dry and clean. Place the cuvette in the measurement chamber, secure the lid, and start the measurement by pressing the Read button. The device will automatically measure after the countdown.

How to calibrate the turbidimeter using the T-CAL standards?

Place the <0.1 NTU standard in the measurement chamber, secure the lid, and press Read. The device will automatically measure after the countdown. Repeat this process for 20 NTU, 200 NTU, and 800 NTU standards, ensuring each is properly prepared and placed in the chamber.

What are the storage conditions for T-CAL standards?

Store T-CAL standards between 5°C and 25°C, avoiding temperatures above 35°C. Keep them away from direct sunlight and allow them to acclimate to the ambient temperature of the turbidimeter before use. Properly stored standards have a shelf life of at least 12 months.

What is the CE conformity declaration for the TB 210 IR?

The TB 210 IR complies with the requirements of the product family standard DIN EN 61326-1:2006. It meets the basic immunity test requirements and emissions standards for class B equipment, as specified in the CE declaration of conformity.

How to handle the T-CAL standards during measurement?

Always use the original cuvettes for T-CAL standards. Ensure the standards are at the same temperature as the turbidimeter before use. Avoid direct sunlight and ensure the cuvette is clean and dry before placing it in the measurement chamber.