Product description:
Frequently Asked Questions
What is the pressure range of the CPP1200-X hydraulic comparison test pump?
The CPP1200-X hydraulic comparison test pump has a pressure range of 0 ... 1,200 bar.
What types of pressure transmission media are compatible with the CPP1200-X?
The CPP1200-X is compatible with hydraulic fluids from mineral oil or clean water, free of calcium carbonate.
What is the capacity of the reservoir in the CPP1200-X?
The reservoir of the CPP1200-X has a capacity of 170 cm³.
What are the dimensions of the instrument base for the CPP1200-X?
The instrument base of the CPP1200-X has dimensions of 401 x 397 x 155 mm (W x D x H).
What is the weight of the CPP1200-X without the storage case?
The CPP1200-X weighs 13.5 kg without the storage case.
What materials are used for wetted parts in the CPP1200-X?
The wetted parts of the CPP1200-X are made of austenitic stainless steel, high tensile brass, and nitrile rubber.
What is the permissible operating temperature range for the CPP1200-X?
The permissible operating temperature range for the CPP1200-X is 0 ... 40 °C.
What is the displacement of the piston in the spindle pump of the CPP1200-X?
The displacement of the piston in the spindle pump of the CPP1200-X is approximately 10 cm³.
What is the recommended pressure medium when using water in the CPP1200-X for pressures over 700 bar?
When using water as the pressure medium in the CPP1200-X for pressures over 700 bar, it is recommended to use a mixture of 90 % water and 10 % glycerine by volume.
What is the CE conformity directive for the CPP1200-X?
The CPP1200-X conforms to the Pressure Equipment Directive 97/23/EC (PS > 1,000 bar; Module A, pressure accessory).