Product description:
Frequently Asked Questions
How to set the safety temperature limit?
To set the safety temperature limit, turn the adjustable safety circuit screw slowly counterclockwise until the heating function turns off and the display shows Er 25. The safety temperature must always be at least 25 °C below the flash point of the medium being processed. The maximum heating plate temperature must be at least 10 °C below the set safety temperature limit.
How to set the temperature control mode?
The user can choose between two temperature control modes using an external PT1000 temperature sensor: PI-Modus for stable results with minimal overshoot, and 2P-Modus for faster heating with more overshoot. To switch modes, set the desired temperature to 2 °C, turn off the device, hold the temperature knob, then turn it back on. The display will show 2P if the 2P mode is active.
What are the safety precautions when using the device?
Ensure the device is placed on a stable, dry, and fireproof surface. Avoid touching the housing or heating plate, as they can reach temperatures over 340 °C. Keep the power cable away from the heating plate. Use personal protective equipment appropriate for the medium being processed. Ensure the device is not used in explosive atmospheres, with hazardous substances, or underwater.
How to operate the device?
Before starting, remove the protective film from the heating plate. Turn the power switch to ON, connect the power cable, and set the desired speed and temperature using the control knobs. The display will show the set values. For heating, press the temperature knob to activate the heating function. The display will alternate between the set and actual temperatures.
What should I do if the device overheats?
If the device overheats, turn it off and let it cool down. Ensure the device is placed on a stable, non-slip surface. Avoid overfilling the container and ensure the container is securely positioned. If the temperature exceeds the safety limit, the device will automatically shut off the heating function and display Er 25.
How to clean the device?
To clean the device, unplug it and use only IKA® recommended cleaning agents. Avoid letting moisture enter the device. For specific cleaning solutions, use isopropyl alcohol or surfactant-containing water depending on the medium being cleaned. Always wear protective gloves when cleaning the device.
What are the error codes and how to resolve them?
Error codes like E3 (overheating), E4 (motor blocked), and E5 (no temperature increase) indicate issues with the device. To resolve these, turn off the device, check for blockages, ensure the temperature sensor is properly immersed, reduce the volume of the medium, or replace the container with a metal pot. For E25 and E26, verify the temperature sensor connections and ensure the safety temperature limit is set correctly.
How to maintain the device?
The device requires no maintenance. Ensure it is used only for its intended purpose and avoid any modifications to the device or circuit board. Regularly inspect the device and accessories for damage before use. Replace any damaged parts immediately to ensure safe operation.
What accessories are available for the device?
Available accessories include stirrers of various sizes, a stirrer remover, a stand rod, a support plate, a holding rod, cross couplings, and contact thermometers. These accessories are designed to enhance the device's functionality and ensure safe and efficient operation.
How to set the time limit for error 5?
To set the time limit for error 5, set the desired temperature to 5 °C, turn off the device, hold the temperature knob, then turn it back on. Use the speed knob to select the time limit between 0 and 30 minutes in 1-minute increments. The display will show the selected time limit, and the device will automatically shut off if the temperature does not increase within this period.
Related products from IKA
C 7030
User manual 3 pages
topolino mobil
User manual 62 pages
topolino mobil
User manual 68 pages
C-MAG HS 7 digital
User manual 28 pages
C-MAG HS 7 digital
User manual 34 pages
C-MAG HS 7 digital
User manual 48 pages
C-MAG HS 7 digital
User manual 60 pages
C-MAG HS 7 digital
User manual 66 pages
C-MAG HS 7 digital
User manual 273 pages
RCT basic
User manual 120 pages
RCT basic
User manual 68 pages
Mini MR standard
User manual 40 pages
Mini MR standard
User manual 40 pages
C-MAG HP4
User manual 28 pages
C-MAG HP4
User manual 34 pages
C-MAG HP4
User manual 60 pages
C-MAG HP4
User manual 66 pages
S 25 KD-LR-18 G
User manual 7 pages
S 25 KD-25 G-ST
User manual 7 pages
S 25 KD-LR-25 G
User manual 7 pages