Product description:
Frequently Asked Questions
How do I change the display from meters to feet?
During calibration, press the MODE key to switch the displayed length from meters to feet. Once the desired cable length or NVP is shown, press the ON/OFF key to exit calibration mode and save the new calibration factor.
What is the purpose of the NVP percentage?
NVP (Nominal Velocity of Propagation) measures how fast a signal travels down a cable compared to the speed of light. It is represented as a percentage of the speed of light and must be set correctly for accurate length tests.
How do I set the NVP percentage?
In calibration mode, use the ▲▼ keys to adjust the displayed NVP until the correct value is shown. The cable length will automatically update based on the new NVP setting.
What happens if the cable length is less than 50 feet during calibration?
Cables shorter than 50 feet (15 meters) will display a FAULT message during calibration. Calibration requires cables that are at least 50 feet long to ensure accurate results.
How do I enter calibration mode?
Turn off the MICROSCANNER, then hold the MODE key while pressing the ON key to start calibration mode. Use this mode to calibrate cable lengths over 50 feet up to 1500 feet.
What type of battery does the MICROSCANNER use?
The MICROSCANNER requires a 9 Volt Alkaline battery. A low battery icon will appear on the screen when the battery level is low, which may affect test accuracy.
Can I use a 9 Volt Carbon Battery with the MICROSCANNER?
No, the MICROSCANNER will not function properly with a 9 Volt Carbon Battery. Only Alkaline batteries are recommended for reliable operation.
How do I check the hardware and software versions?
Turn off the MICROSCANNER, then hold the ▲ and MODE keys while pressing the ON key to display the hardware and software versions.
What should I do if I have technical questions?
Contact Fluke Networks' Technical Support via phone, fax, or email. Provide your hardware and software version numbers before calling. Support numbers are available for the US, Canada, Europe, and other regions.
What is the function of the high voltage protection feature?
The MICROSCANNER is designed to withstand normal telephony voltage conditions, such as 48 VDC or 24 VAC. Testing should not be performed if hazardous conditions are present on the inputs.