Product description:

The WIKA CPB3500 is a high-precision pressure testing device designed for use in industrial and laboratory environments. It features a robust construction and a digital display for accurate pressure measurement. The device supports a wide range of pressure ranges and is suitable for both static and dynamic pressure testing. It is equipped with various connection options to ensure compatibility with different types of test systems and components. The CPB3500 is ideal for applications requiring reliable and repeatable pressure measurements, such as in quality control, process engineering, and research.

Frequently Asked Questions

What is the purpose of the CPB3500 pneumatic dead-weight tester?

The CPB3500 pneumatic dead-weight tester is designed for calibrating electronic or mechanical pressure measuring instruments. It provides precise pneumatic pressure ranges from -1 ... 120 bar [-14.5 ... 1,600 lb/in²] using a piston-cylinder system and mass sets.

What are the key components of the CPB3500?

The CPB3500 includes an instrument base with an ABS cover, a piston-cylinder system, mass sets, VG22 mineral oil, a standard tool set, and an integrated precision dual-area spindle pump for pressure generation.

What safety precautions should be taken when using the CPB3500?

The CPB3500 should only be used in non-hazardous areas with clean, dry gas. Avoid using it with liquid media or for calibrating oxygen gauges. Ensure the instrument is not exposed to oil or hydrocarbons, as this could lead to explosions.

What are the recommended storage conditions for the CPB3500?

Store the CPB3500 in a location with a temperature range of -10 ... +50 °C [14 ... 122 °F] and relative humidity between 35 ... 85% (for the base and mass set) or 35 ... 65% (for the piston-cylinder system). Avoid exposure to direct sunlight, mechanical vibrations, and corrosive gases.

How should the CPB3500 be transported?

Transport the CPB3500 carefully to avoid damage. Do not remove the packaging until just before installation. If transporting from a cold to a warm environment, allow the instrument to acclimate to room temperature before use to prevent condensation.

What is the maximum pressure the CPB3500 can handle?

The CPB3500 can handle a maximum pressure of 120 bar [1,600 lb/in²]. Exceeding this pressure may damage the piston-cylinder system and compromise the accuracy of the calibration.

What is the function of the volume adjuster in the CPB3500?

The volume adjuster in the CPB3500 allows for precise control of the pressure during calibration. It ensures accurate comparative testing by adjusting the volume of the system to match the required pressure settings.

What is the role of the integrated dual-area spindle pump in the CPB3500?

The integrated dual-area spindle pump in the CPB3500 is used to generate and regulate the pressure. It ensures that the system reaches equilibrium, allowing for accurate calibration or adjustment of the test item.

What are the calibration requirements for the CPB3500?

The CPB3500 should be calibrated according to international standards. Regular calibration is necessary to maintain accuracy, especially after any maintenance or if the instrument has been used in different environments. Factory calibrations are performed in accordance with UKAS standards.

What should be done if the CPB3500 is to be used for oxygen-related applications?

The CPB3500 is not suitable for use with oxygen. If oxygen-related applications are required, the instrument must be modified to be oil-free, and this should be clearly documented. Always ensure that no oil or hydrocarbons are present to prevent the risk of explosion.