Product description:

The WIKA CPP120-X is a pressure testing device designed for precise measurement and calibration of pressure instruments. It features a robust construction and is suitable for use in industrial and laboratory environments. The device supports a wide range of pressure ranges and is equipped with a digital display for accurate readings. It is designed for ease of use and ensures reliable performance in various applications.

Frequently Asked Questions

What is the maximum pressure capacity of the CPP120-X?

The CPP120-X pneumatic comparison test pump is designed to provide pressure up to 120 bar [1,740 psi]. This makes it suitable for precise comparative testing of pressure measuring instruments.

What components are included in the scope of delivery?

The scope of delivery includes the instrument base, two fine metering valves for pressure inlet and outlet, a volume adjuster, two test item connections with G ½ female thread, an external pressure/vacuum connection, and adapter sets for different thread types. A tool and maintenance set are also provided.

What safety precautions should be taken when using the CPP120-X?

Always read and follow the safety instructions in the operating manual. Avoid using the pump in hazardous areas, ensure proper personal protective equipment is worn, and only trained personnel should operate the device. The pump requires a regulated supply of clean, dry gas.

How is the test item connected to the CPP120-X?

The test item is connected via G ½ female thread connections on the instrument. These connections allow for the attachment of the test item and a reference pressure measuring instrument, enabling precise comparative testing.

What is the purpose of the volume adjuster?

The volume adjuster allows for precise pressure adjustment during testing. It ensures accurate comparative measurements by fine-tuning the pressure applied to the test item and reference instrument.

What type of gas is recommended for operation?

The pump requires a regulated supply of clean, dry gas. Nitrogen is recommended as it is typically dry and clean. Compressed air can also be used if it is filtered and dried to remove oil and moisture.

What are the safety symbols and their meanings?

The manual includes warning and caution symbols. A warning symbol indicates a potentially dangerous situation that could result in serious injury or death. A caution symbol indicates a situation that could lead to minor injuries or property damage. Information symbols provide useful tips and recommendations.

Who should operate the CPP120-X?

Only trained and qualified personnel should operate the CPP120-X. These individuals must have the necessary technical training and understanding of measurement and control technology to ensure safe and accurate use.

What are the environmental requirements for operation?

The pump requires a stable environment with proper ventilation and a regulated supply of clean, dry gas. It should be used in a location that meets safety and operational standards to ensure accurate and reliable testing.

How should the pump be maintained?

Regular maintenance includes checking and replacing seals, cleaning gauges, and inspecting valves. The pump should be serviced by authorized personnel to ensure continued safe and accurate operation.