WIKA CPB5800 (76 pages)
Tags
Languages available:
No exact matches found. Please change your query
Need help?
Do you have a question about the product that is not answered in the manual?
- Receive updates on solutions
- Get answers to your question
Product description:
Frequently Asked Questions
How is the pressure balance CPB5800 calibrated?
The CPB5800 pressure balance is calibrated to standard reference conditions, which are typically 20°C and standard gravity of 9.80665 m/s². This ensures accurate measurements under normal operating conditions.
What are the main components of the CPB5800 pressure balance?
The CPB5800 consists of a precision-manufactured piston-cylinder system, a mass set, an integrated priming pump, and a spindle pump. These components work together to generate and measure pressure accurately.
What are the different types of piston-cylinder systems available for the CPB5800?
The CPB5800 offers two types of piston-cylinder systems: single-range systems for 120 bar (1,600 psi) and 300 bar (4,000 psi), and dual-range systems for 700 bar, 1,200 bar, and 1,400 bar (10,000 psi, 16,000 psi, and 20,000 psi).
How does the CPB5800 handle local variations in gravity?
Local variations in gravity can affect measurement accuracy. To account for this, the CPB5800 can be calibrated to the local gravity value, and a correction formula is applied to adjust the measured pressure accordingly.
What is the impact of temperature on the CPB5800's performance?
Temperature affects the effective area of the piston-cylinder system. A temperature coefficient (TK) is used to correct for deviations from the standard reference temperature of 20°C, ensuring accurate pressure readings.
How does the CPB5800 handle ambient conditions?
Ambient conditions such as air pressure, room temperature, and relative humidity can influence measurement accuracy. The CPB5800 accounts for these factors by adjusting for air density and humidity effects on the mass set.
What is the recommended recalibration interval for the CPB5800?
The CPB5800 should be recalibrated every 2 to 5 years, depending on usage conditions. This ensures long-term stability and accuracy of the pressure balance.
How is the CPB5800 used for calibration?
The CPB5800 is used for calibrating electronic or mechanical pressure measuring instruments. It generates precise pressure values using a piston-cylinder system and calibrated masses, allowing for accurate comparisons and adjustments.
What are the safety instructions for operating the CPB5800?
The CPB5800 should only be operated by trained personnel. It must be handled with care to avoid damage, and any unauthorized modifications or repairs can void the warranty and compromise safety.
What are the maintenance requirements for the CPB5800?
Regular maintenance includes cleaning the piston-cylinder system, checking the mass set, and replacing consumable parts as needed. The hydraulic pressure medium should be changed periodically, and the system should be vented properly after filling.