Elcometer 130 (126 pages)
Tags
Languages available:
No exact matches found. Please change your query
Need help?
Do you have a question about the product that is not answered in the manual?
- Receive updates on solutions
- Get answers to your question
Product description:
Frequently Asked Questions
How to switch the gauge on?
Press and hold the N/ button to switch the gauge on. Ensure the filter paper size is set as required, see section 3.4. Place the gauge on a clean, dry surface. Apply the precise amount of high purity water required, dependent on the filter paper size. Use tweezers to remove a filter paper from the pack.
How to position the filter paper?
For half-size filter paper, position it on the bottom half of the circle using the markers on the measurement electrodes as a guide. For quarter-size filter paper, position it on the bottom right quadrant of the circle using the markers on the measurement electrodes as a guide.
How to close the lid?
Close the lid ensuring that the magnetic catch is fully engaged. The reading will automatically be displayed on screen and stored into memory together with the filter paper size, temperature, date and time (when in batch mode).
How to remove the filter paper?
After the 2-minute test time, remove the filter paper from the test surface and place it on the measurement electrodes. Ensure the filter paper is positioned correctly on the measurement electrodes. If manual setup is used, the correct filter paper size is used. No manual calculations are required by the user.
How to clean the electrodes?
Clean the electrodes between tests using pure water and a sensor wipe (supplied). Thorough cleaning of the electrodes between tests may contaminate subsequent tests and result in inaccurate readings. The electrodes are coated to prevent corrosion and oxidation, prolonging the life and accuracy of the gauge. Do not use abrasive materials as this will damage the electrodes.
How to calibrate the offset function?
To measure soluble salts, high purity water should be used. Non-pure water, with a maximum conductivity of 2.5 /cm (equivalent to 2 g/cm² or 2 mg/m²), can be used by setting a calibration offset on the cometer 0. The calibration offset function is used to set the offset value.
How to set the measurement units?
The cometer 0 provides readings in g/cm², the measurement units for surface cleanliness, conductivity, and other parameters. To select the measurement units, press enu/ etu / units. The cometer 0 has a choice of g/cm² or mg/m², /cm, m /cm, m or C/
How to select the filter paper size?
The cometer 0 can be set to automatically detect the filter paper size in use or this can be manually selected by the user. To select the filter paper size, press enu/ etu / filter paper size. For half-size filter paper, take the standard filter paper in hand and cut it into two along the fold line. For quarter-size filter paper, take the standard filter paper into quarters and cut into quadrants along the fold lines.
How to use the 2-minute timer?
The wetted filter paper should be placed on the surface under test for 2 minutes. The cometer 0 has an optional timer for this purpose. To enable the timer, press enu/ etu / timer. When enabled, the timer is activated as required. The filter paper should be discarded and the sample retested if the filter paper is placed on the substrate for more than 2 minutes.
How to use temperature compensation?
The temperature of the substrate will affect the reading. As the cometer 0 has been calibrated at 25°C, to provide an accurate measurement of soluble salts, the reading must be adjusted to take into consideration a temperature variance. The automatic temperature compensation function on the cometer 0 measures the temperature of the wet filter paper when placed on the measurement electrodes and using this value, adjusts the reading accordingly. To activate, press enu/ etu / temperature compensation.
Related products from Elcometer
500
User manual 252 pages
415
User manual 112 pages
138
User manual 32 pages
138
User manual 32 pages
138
User manual 48 pages
103
User manual 28 pages
456
User manual 112 pages
456
User manual 112 pages
456
User manual 72 pages
456
User manual 72 pages
456
User manual 28 pages
116C
User manual 33 pages
224
User manual 70 pages
2352
User manual 60 pages
2351
User manual 60 pages
2350
User manual 60 pages
2354
User manual 60 pages
2353
User manual 60 pages
550
User manual 80 pages
MTG2
User manual 140 pages