Megger BM80/2 (89 pages)
Tags
Languages available:
No exact matches found. Please change your query
Need help?
Do you have a question about the product that is not answered in the manual?
- Receive updates on solutions
- Get answers to your question
Product description:
Frequently Asked Questions
What is the maximum insulation test voltage available?
The BM80/2 Series offers insulation test ranges of 250 V, 500 V, 1000 V, and 100 V. The 1000 V range is suitable for testing electrical installations in compliance with BS7671, IEC364, and HD384 standards.
How does the instrument handle live circuits during testing?
The instrument automatically inhibits testing if an external voltage greater than 55 V is detected on insulation test ranges or greater than 25 V on other ranges. A warning is displayed, and the bleeper sounds intermittently to alert the user.
What is the purpose of the Polarization Index (PI) test?
The Polarization Index test measures the Dielectric Absorption Ratio by comparing resistance values after 1 minute and 10 minutes. It helps assess insulation quality and is useful for identifying potential faults in electrical systems.
How does the continuity test work on the BM80/2 Series?
Continuity tests are activated automatically when the probes make contact. The test operates without pressing the TEST button, and the reading is held for a few seconds after the test leads are disconnected. The continuity bleeper sounds for low resistance values, indicating a good connection.
How do you zero the test lead resistance?
To zero the test lead resistance, short the test leads together, wait for a stable reading, and press the TEST button. The zero offset symbol appears to indicate the adjustment is complete. This ensures accurate continuity measurements.
What is the battery check function used for?
The battery check function allows you to monitor the remaining battery life at any time. It is particularly useful before going on-site to ensure sufficient power for the day’s work. A low battery symbol appears on the display when power is nearly exhausted.
What is the auto-shut-off feature and how does it work?
The auto-shut-off feature conserves battery life by turning off the instrument after 12 minutes of inactivity on insulation test ranges and 5 minutes on other ranges. This can be adjusted to 60 minutes for non-insulation test ranges. The instrument must be switched to OFF when not in use to prevent unnecessary power consumption.
What are the safety precautions when using the instrument?
The instrument must be used only by trained and competent personnel. The circuit under test must be de-energized and isolated before connections are made. Fused test leads are recommended for high-energy systems to prevent electric shock. Always ensure capacitive circuits are fully discharged before disconnecting test leads.
What is the accuracy of the BM80/2 Series instruments?
The BM80/2 Series instruments have a basic accuracy of ±2% at 20°C. They are designed to IEC1010-1 standards and are protected against connection to a 440 V Category III supply, ensuring reliable performance in various environments.
How do you perform a voltage test?
To perform a voltage test, set the selector switch to the V position, connect the test leads, and the instrument will display the measured a.c. or d.c. voltage. The display can be toggled between different voltage ranges, and the instrument will indicate negative d.c. with the '-dc' symbol.
Related products from Megger
OTS80AF
User manual 12 pages
OTS60AF
User manual 12 pages
OTS 60PB
User manual 12 pages
OTS 60PB
User manual 56 pages
WM6
User manual 22 pages
SCITS100
User manual 35 pages
S1-5001
User manual 58 pages
BM21
User manual 71 pages
DET5/4D
User manual 72 pages
DET5/4R
User manual 72 pages
DET10C
User manual 79 pages
DET20C
User manual 79 pages
BM25
User manual 88 pages
S1-5005
User manual 88 pages
OTS 100AF/2
User manual 104 pages
OTS 80AF/2
User manual 104 pages
LCB2000/2
User manual 136 pages