Megger OTS 80AF/2 (104 pages)
Tags
Languages available:
No exact matches found. Please change your query
Need help?
Do you have a question about the product that is not answered in the manual?
- Receive updates on solutions
- Get answers to your question
Product description:
Frequently Asked Questions
How often should the continuity check between the EMC/discharge barrier and earth point be performed?
A routine continuity check between the EMC/discharge barrier and the earth point should be made weekly to ensure proper functionality and safety.
What type of oil test vessel must be used with the OTS series?
The correct type of oil test vessel must be used, and it must be correctly positioned on the support horns before carrying out any testing to ensure accurate results.
What should be done if the test chamber cover is cracked or damaged?
If the test chamber cover is cracked or damaged in any way, the test set must not be used and should be sent for repair to an authorised agent to avoid safety risks.
What is the maximum output voltage for the OTS100AF/2?
The OTS100AF/2 has a maximum output voltage of 100 kV, with a range of 50 kV - 0 - 50 kV, and operates at a high voltage test frequency of 61.8 Hz.
What safety feature prevents high voltage from being applied to the test chamber?
The hinged cover to the test chamber is fastened by a trapped key interlock switch, which is a safety feature that prevents high voltage from being applied unless the interlock is closed.
What should be done if the test set is exposed to severe environmental changes?
If the oil test set has been exposed to severe environmental changes, such as humidity or temperature variations, it should be allowed to acclimatise to its new conditions for at least 12 hours before use.
What is the recommended method for cleaning the test vessel?
The test vessel should be wiped clean with a dampened cloth using Isopropyl Alcohol (IPA) to ensure it is free from contaminants before testing.
What is the purpose of the trapped key interlock switch?
The trapped key interlock switch is a safety feature that ensures high voltage cannot be applied to the test chamber unless the interlock is closed, preventing accidental exposure to high voltage.
What type of fuses are required for the OTS series?
Replacement fuses must be of the correct type and rating, specifically ceramic fuses conforming to IEC 127 standards, as glass fuses are not suitable due to their low breaking capacity.
What should be done if the test set is suspected to have impaired protection?
If the test set's protection is suspected to be impaired, it must not be used and should be taken out of service and returned to the manufacturer or an authorised agent for repair.
Related products from Megger
OTS80AF
User manual 12 pages
OTS60AF
User manual 12 pages
OTS 60PB
User manual 12 pages
OTS 60PB
User manual 56 pages
WM6
User manual 22 pages
SCITS100
User manual 35 pages
S1-5001
User manual 58 pages
BM21
User manual 71 pages
DET5/4D
User manual 72 pages
DET5/4R
User manual 72 pages
DET10C
User manual 79 pages
DET20C
User manual 79 pages
BM25
User manual 88 pages
S1-5005
User manual 88 pages
BM80/2
User manual 89 pages
OTS 100AF/2
User manual 104 pages
LCB2000/2
User manual 136 pages