Product description:

The Dräger X-am 5000 is a multi-gas detection device designed for use in industrial and environmental applications. It offers high accuracy and reliability in detecting a wide range of gases, including toxic and combustible gases. The device features a robust construction suitable for harsh environments and includes a user-friendly interface for easy operation. It supports various gas configurations and can be customized according to specific application requirements. The X-am 5000 is equipped with advanced sensor technology and provides real-time data for immediate response to potential hazards.

Frequently Asked Questions

How often should the sensitivity be checked?

The sensitivity must be checked daily before the first use with a known concentration of the gas being measured, corresponding to 25 to 50 percent of the concentration end value. The accuracy must be between 0 and +20 percent of the actual value. Accuracy can be corrected through calibration.

What gases can the X-am 5000 measure?

The X-am 5000 is a portable gas meter for continuous monitoring of the concentration of multiple gases in ambient air at the workplace and in explosive areas. It can independently measure up to five gases according to the installed Dräger sensors.

What are the operating conditions for the X-am 5000 in explosive areas?

The device is designed for use in explosive areas classified as Zone 0, Zone 1, or Zone 2, or in mines affected by methane gas. It is intended for use within a temperature range of -20 °C to +50 °C, and for areas where gases of explosion classes IIA, IIB, or IIC and temperature class T3 or T4 (depending on the battery and batteries) may be present. When used in mines, the device can only be used in areas where there is a low risk of mechanical influences.

What are the alarm configurations for the X-am 5000?

Alarms A1 and A2 have configurable self-holding and acknowledgment features using the Dräger CC-Vision PC software. Alarm A1 is set to trigger with a rising slope and includes a pre-alarm, while Alarm A2 is self-holding and acts as the main alarm with a falling slope. Both alarms are non-acknowledgable unless configured otherwise.

What are the default settings for the X-am 5000?

The default settings for the X-am 5000 include enabling the bump test mode, fresh air calibration, and operational signals. The device is set to allow shutdown, with a UEG factor of 4.4 Vol.-% for CH4, a STEL function inactive with a 15-minute averaging time, and a TWA function inactive with an 8-hour averaging time. All alarms are non-acknowledgable unless configured otherwise.

How do you check the device parameters?

To ensure the values have been correctly transferred to the gas meter, select the 'Data from X-am 1/2/5x00' button in the Dräger CC-Vision software and check the parameters. All alarm thresholds A1 and A2, as well as TWA and STEL for all toxic gases, are displayed sequentially. During sensor warm-up, the measurement value display and the warning symbol will blink, and no alarms are triggered during this phase.

What should be done before turning on the device?

Before using the device for the first time, insert a charged NiMH power unit T4 or Dräger-approved batteries, as described in the 'Changing Batteries' section on page 9. Ensure the device is ready for operation. Hold the [OK] and [+] buttons simultaneously until the countdown '3. 2. 1' on the display completes. Before shutting down, the optical, acoustic, and vibration alarms will briefly activate.

What should be checked before entering the work area?

Before conducting safety-critical measurements, verify the calibration using a bump test (Bump Test), adjust if necessary, and check all alarm elements. If national regulations apply, perform the bump test according to these regulations. Incorrect calibration can lead to false measurement results, which may result in serious health damage. Only use power units of type ABT 01xx, HBT 00xx, or HBT 01xx. Avoid using the device in oxygen-enriched atmospheres (>21 Vol.-% O2) as it may compromise explosion protection.

What are the safety precautions when using the device in explosive environments?

When using the device in explosive environments, avoid exposing the CatEx sensor to catalytic poison components in the measurement gas, such as volatile silicon, sulfur, heavy metal compounds, or halogenated hydrocarbons. These can damage the sensor. If the CatEx sensor cannot be recalibrated to the target concentration, replace it. Avoid using the device in oxygen-deficient atmospheres (<8 Vol.-% O2) as it may lead to unreliable readings. Do not use the device in oxygen-enriched atmospheres (>21 Vol.-% O2) as it may compromise explosion protection. High values outside the display range may indicate an explosive concentration.

What should be done during device operation?

During operation, ensure the gas inlet opening is not obstructed or dirty. Monitor the display for any warning symbols or disturbance indications. If a warning is displayed and does not automatically clear during operation, the device must be serviced after use. The device is not ready for measurement and must be serviced if it shows any issues.