Product description:

The Dräger X-am 5600 is a multi-gas detection device designed for use in industrial and occupational safety environments. It offers real-time monitoring of various gases, including toxic and combustible gases, as well as oxygen levels. The device features a robust construction suitable for harsh environments and is equipped with a high-resolution display for clear and immediate readings. It supports multiple gas detection ranges and can be configured for different application scenarios. The X-am 5600 is ideal for use in confined spaces, emergency response, and routine safety inspections.

Frequently Asked Questions

How to power on the device?

To turn on the device, press and hold the [ OK ]-button for approximately 3 seconds until the countdown displayed on the screen '3 . 2 . 1' completes. This activates all display segments, the optical, acoustic, and vibration alarms to check the proper function. The software version is shown, and the device performs a self-test. The next sensor to be calibrated and the remaining days until the next calibration are displayed, along with the time until the end of the bump test interval.

How to power off the device?

To turn off the device, press and hold both the [ OK ]-button and [ + ]-button until the countdown displayed on the screen '3 . 2 . 1' completes. Before shutting down, the optical, acoustic, and vibration alarms are briefly activated.

What is the purpose of the device?

The Dräger X-am® 5600 is a portable gas measurement device for continuous monitoring of the concentration of multiple gases in ambient air at the workplace and in explosive areas. It independently measures up to 6 gases according to the installed Dräger sensors.

What are the approved usage conditions?

The device is intended for use in explosive areas classified as Zone 0, Zone 1, or Zone 2, or in mines affected by methane gas. It is designed for use within a temperature range of -20 °C to +40 °C or +50 °C, and for areas where gases of Explosions Class IIA, IIB, or IIC and Temperature Class T3 or T4 (depending on the battery and batteries) may be present. When used in mines, the device can only be used in areas where there is a low risk of mechanical influence.

What are the safety precautions when using the device?

To reduce the risk of ignition of flammable or explosive atmospheres, the following precautions and warning notes must be observed: Do not mix new batteries with used ones or batteries from different manufacturers. Disconnect the power unit from the device before maintenance. The replacement of components may affect the intrinsic safety. The device should not be used in atmospheres enriched with oxygen (>21% O2).

How to perform a bump test?

A bump test is allowed and is used to verify the calibration of the device. The bump test interval is set to 4.4 Vol.-% (4.4 Vol.-% correspond to 100 %UEG) for the first test and 4.0 Vol.-% (4.0 Vol.-% correspond to 100 %UEG) for the second test. The device will display the remaining days until the next calibration and the time until the end of the bump test interval.

What are the alarm functions?

The device has two alarm functions: Alarm A1 and Alarm A2. Alarm A1 is triggered by the O2 sensor and has three modes: non-acknowledgeable, self-holding, and rising edge. Alarm A2 is non-acknowledgeable, self-holding, and falling edge. Both alarms can be configured using the PC software Dräger CC-Vision.

How to configure the device?

Only trained and qualified personnel may change the device configuration. To configure the device individually with a standard configuration, connect it via USB-infrared cable (Order No. 83 17 409) or the E-Cal system to a PC. Configuration is performed using the PC software Dräger CC-Vision, which can be downloaded for free from the X-am 5600 product page at www.draeger.com. The current settings can be checked and changed using this software.

What are the operating parameters?

The device allows changes to several operating parameters, including the operation signal LED, operation signal horn, shutdown mode, shift length (TWA), short-term exposure duration (STEL), and alarm thresholds. These parameters can be adjusted within specified ranges, and the settings can be verified through the software.

What are the sensor settings?

For each sensor, you can adjust the alarm thresholds (A1 and A2), the evaluation type (inactive, TWA, STEL, TWA+STEL), and the alarm thresholds for STEL and TWA. These settings are applicable only if the sensor is designed for that purpose. The sensor evaluation type determines how the exposure values are calculated.

Send answer